1 – 10 k high Precision transportable Setup to calibrate Multifunction Electrical instruments
نویسنده
چکیده
A temperature controlled 1 –10 kΩ standard Resistors transportable setup was developed at National Institute of Metrological Research, (INRIM) for the calibration and adjustment of multifunction electrical instruments. The two Standards consist respectively of two 10 and 100 k parallel connected resistors nets inserted in a temperature controlled aluminium box. Novelty of the realization is the oil insertion of the 1 net with its internal connectors lowering the thermo-electromotive forces (emfs) effects. Short and mid-term stabilities of the setup Standards resulted on the order and in some cases better than other top level 1 and 10 kΩ commercial Standards. The transport effect turning off the setup temperature control did not cause appreciable measurement deviations on the two Standards. The Standards uncertainties meet those requested by DMMs and MFCs manufacturers to calibrate and adjust these instruments. A test to adjust a multifunction calibrator gave satisfactory results.
منابع مشابه
1 Ω – 10 k Ω High Precision Resistance Setup to calibrate Multifunction Electrical instruments
A temperature controlled 1 Ω–10 kΩ Standard resistors setup was developed at National Institute of Metrological Research, (INRIM). The aim of this realization was the involvement of the setup resistors in the traceability transfer process to high accuracy multifunction electrical instruments used in Secondary Electrical Calibration Laboratories or even their use as primary Standards in high lev...
متن کامل1 Ω–10 kΩ high accuracy standard setup for calibration of multifunction electrical instruments
A temperature controlled 1 Ω and 10 kΩ standard resistors setup has been developed at National Institute of Metrological Research, (INRIM), to transfer the traceability to high accuracy multifunction electrical instruments used in Secondary calibration laboratories. These two standards are formed respectively by ten in parallel connected 10 Ω and 100 kΩ resistors inserted into a temperature con...
متن کاملAccuracy and precision of two laser dendrometers
Two commercial laser dendrometers were tested under controlled and field conditions, and contrasted with alternative instruments. Testing focused on height measurement, but also considered distance and remote diameter measurements. Both laser instruments gave very precise estimates, but showed some bias. Users of these and other `high-tech' instruments are reminded that precision is not synonym...
متن کاملNovel Axial Flux Brushless Resolver Analysis and Optimization Using 3D Finite Element and d-q Model Method
Resolvers are widely used in electric driven systems especially in high precision servomechanisms. Both encapsulated and pancake resolvers suffer from a major drawback: static eccentricity (SE). This drawback causes a significant increase in resolver output position error (RPE) which could not be corrected electronically. To reduce RPE, this paper proposes a novel structure with axial flux. Pr...
متن کاملLow Flow Liquid Calibration Setup
This article describes a primary calibration setup, and its uncertainty, for low flow liquid calibrations at Bronkhorst High-Tech. It will be used to calibrate reference flow meters from 1 to 200 g/h. By setting up an uncertainty budget for this setup, the calibration of the instruments can be compared to that of NMI’s (National Metrology Institutes). The uncertainty budget consists of mass, ti...
متن کامل