Validation of an ultrasensitive digital droplet PCR assay for HIV-2 plasma RNA quantification

نویسندگان

  • Jean Ruelle
  • Vasilieios Yfantis
  • Armelle Duquenne
  • Patrick Goubau
چکیده

INTRODUCTION Low or undetectable plasma viral load (VL) using current qPCR assays is common for HIV-2 patients. Digital PCR is an emerging technology enabling more precision and reproducibility than qPCR at low DNA/RNA copy numbers. Available data related to digital droplet PCR (ddPCR, Bio-Rad) underscore issues linked to the threshold definition of positivity, coupled to the specificity of low copy results (1). MATERIALS AND METHODS A RT-PCR protocol was set up using the One-Step RT-ddPCR Kit for Probes on the QX200 platform (Bio-Rad, Hercules, CA) in an accredited environment (ISO15189:2012 norm). Parameters tested were in line with the digital MIQE guidelines (2). Inter-run coefficient of variation (CV) was established using synthetic RNA controls diluted in HIV-negative plasma. The ddPCR assay was compared to a qRT-PCR previously used in routine (LOQ 50 cop/mL (3)) using 46 clinical samples and the NIBSC international HIV-2 RNA standard. RESULTS The optimal PCR efficiency and the best separation between positive and negative droplets were obtained with a mixture containing 0.5 mM manganese acetate, 700 nM primers and 250 nM of the 5'FAM-probe. Using a manual threshold to define positivity, 7.74% of negative controls (n=168) were scored as positive due to one positive droplet. The presence of two positive droplets or more was not observed for negative controls. Serial dilutions of a positive control showed excellent linearity (R2=0.999) and enabled us to define a limit of quantification of two positives droplets, which corresponds to 0.14 copies/μL in the reaction mixture and to seven copies per mL of plasma. The inter-run coefficient of variation was 3.37% at a mean value of 4,468 cop/mL, 19.59% at 416 cop/mL and 32.28% at 8 cop/mL. The NIBSC standard of 1,000 IU was quantified 1,400 copies by ddPCR and close to 5,000 copies by qPCR (delta log superior to 0.5). Among 46 clinical samples, 22 were undetectable with both qPCR and ddPCR, 12 were detected with both methods (respective means of 10,612 and 2,224 cop/mL, delta log=0.68) and 12 others were quantified by ddPCR only below 50 cop/mL (mean=16 cop/mL). CONCLUSIONS We validated a ddPCR HIV-2 VL assay that is more sensitive and more reproducible than the qPCR assay used as comparator, with a limit of quantification of 7 cop/mL of plasma. A careful definition of the limit of blank allows the management of false positive droplets, but the variable user-defined positive threshold may be an issue for compliance to the quality norms.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Quantification of HEV RNA by Droplet Digital PCR

The sensitivity of real-time PCR for hepatitis E virus (HEV) RNA quantification differs greatly among techniques. Standardized tools that measure the real quantity of virus are needed. We assessed the performance of a reverse transcription droplet digital PCR (RT-ddPCR) assay that gives absolute quantities of HEV RNA. Analytical and clinical validation was done on HEV genotypes 1, 3 and 4, and ...

متن کامل

تعیین کمی بار ویروسی هپاتیت C با استفاده از روش Real-Time PCR In-House در بیماران آلوده به هپاتیت C در شهرستان خرم آباد

Background : Molecular diagnostic methods are among major tools in management of hepatitis C virus (HCV) in infected patients. Many studies have shown that viral load is associated with stage of infection and response to treatment. Therefore, the evaluation and quantification of viral load is very important. The goal of this study is implementation of inexpensive, yet accurate method for quanti...

متن کامل

Ultrasensitive reverse transcription-PCR assay for quantitation of human immunodeficiency virus type 1 RNA in plasma.

With the recent introduction of combination therapy, human immunodeficiency virus type 1 (HIV-1) RNA levels in plasma have been dramatically reduced, frequently to below the limit of quantitation (400 copies/ml of plasma) of the AMPLICOR HIV-1 MONITOR Test (Roche Diagnostic Systems). To achieve enhanced sensitivity of the AMPLICOR HIV-1 MONITOR Test, a modified specimen preparation procedure th...

متن کامل

Comparison of Droplet Digital PCR and Seminested Real-Time PCR for Quantification of Cell-Associated HIV-1 RNA

Cell-associated (CA) HIV-1 RNA is considered a potential marker for assessment of viral reservoir dynamics and antiretroviral therapy (ART) response in HIV-infected patients. Recent studies employed sensitive seminested real-time quantitative (q)PCR to quantify CA HIV-1 RNA. Digital PCR has been recently described as an alternative PCR-based technique for absolute quantification with higher acc...

متن کامل

Factors Associated With Persistence of Plasma HIV-1 RNA During Long-term Continuously Suppressive Firstline Antiretroviral Therapy

Background Persistence of plasma HIV-1 RNA during seemingly effective antiretroviral thereapy (ART) is incompletely understood. Using an ultrasensitive assay, this cross-sectional study investigated residual plasma HIV-1 RNA in subjects maintained on firstline ART with continuous viral load suppression <50 copies/mL for ≤15 years without recognized viral load blips or treatment interruptions an...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 17  شماره 

صفحات  -

تاریخ انتشار 2014