An evaluation of 2 new devices for nasal high-flow gas therapy.

نویسندگان

  • Jonathan B Waugh
  • Wesley M Granger
چکیده

BACKGROUND The traditional nasal cannula with bubble humidifier is limited to a maximum flow of 6 L/min to minimize the risk of complications. We conducted a bench study of 2 new Food and Drug Administration-approved nasal cannula/humidifier products designed to deliver at flows> 6 L/min. METHODS Using a digital psychrometer we measured the relative humidity and temperature of delivered gas from each device, at 5 L/min increments over the specified functional high-flow range. RESULTS The Salter Labs unit achieved 72.5-78.7% relative humidity (5-15 L/min range) at ambient temperature (21-23 degrees C). The Vapotherm device achieved 99.9% relative humidity at a temperature setting of 37 degrees C (5-40 L/min). CONCLUSIONS Both devices meet minimum humidification standards and offer practical new treatment options. The patient-selection criteria are primarily the severity of the patient's condition and cost.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Heated Humidified High-Flow Nasal Cannula Versus Nasal Continuous Positive Airway Pressure for the Facilitation of Extubation in Preterm Neonates with Respiratory Distress

Background: Heated humidified high-flow nasal cannula (HHHFNC) is gaining popularity as an alternative to nasal continuous positive airway pressure (nCPAP) therapy in the management of preterm neonates with respiratory distress due to ease of administration and patient comfort. However, limited evidence is available addressing its risks and benefits. To study the efficacy and safety of HHHFNC i...

متن کامل

Gas Flow Metering Using the PSO Optimized Interval Type- 2 Fuzzy Neural Network

Orifice flow meter is one of the most common devices in industry which is used for measuring the gas flow. This system includes an orifice plate, temperature and pressure transmitters, and a flow computer. The flow computer is used for collecting information related to temperature, pressure, and their differences under various conditions. Also the flow computer can calculate the flow rate of ga...

متن کامل

Comparison of the Heated Humidified High-flow Nasal Cannula with Nasal Continuous Positive Airway Pressure as Primary Respiratory Support for Preterm Neonates: A Prospective Observational Study

Background: Heated humidified high-flow nasal cannula (HHHFNC) is gaining popularity in the management of respiratory distress in preterm neonates. However, it is not known whether it takes precedence over the gold standard nasal continuous positive airway pressure (NCPAP) in this age group as a primary mode of non-invasive ventilation (NIV). There is limited evidence ...

متن کامل

Humidification performance of two high-flow nasal cannula devices: a bench study.

INTRODUCTION Delivering heated and humidified medical gas at 20-60 L/min, high-flow nasal cannula (HFNC) creates low levels of PEEP and ameliorates respiratory mechanics. It has become a common therapy for patients with respiratory failure. However, independent measurement of heat and humidity during HFNC and comparison of HFNC devices are lacking. METHODS We evaluated 2 HFNC (Airvo 2 and Opt...

متن کامل

High-flow nasal cannula oxygen therapy during hypoxemic respiratory failure.

Oxygen therapy remains the first line intervention in acute hypoxemic respiratory failure. Several medical devices for oxygen therapy, which range from simple nasal cannula to non-rebreathing face masks, have been used in the management of acute hypoxemic respiratory failure.1 The choice of a specific oxygen delivery device is based on the patient’s oxygen requirements in terms of flow and desi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Respiratory care

دوره 49 8  شماره 

صفحات  -

تاریخ انتشار 2004