SU-E-T-535: Preliminary 2D and 3D Gamma Calculation Comparison Using PRESAGE.
نویسندگان
چکیده
PURPOSE To compare gamma calculations from 2D and 3D dosimetry measurements for phantom quality assurance. METHODS An IROC Houston (RPC) head and neck phantom was irradiated with a 9 beam IMRT plan using two inserts: a TLD and film insert and a PRESAGE insert. Both inserts were irradiated 3 times. The film and PRESAGE doses were scaled to the TLD dose and 2D gamma calculations were made in the axial and sagittal planes bisecting the primary target. 3D gamma measurements were taken within the PRESAGE dosimeter volume. Gamma constraints of 3%/3mm distance to agreement (DTA), 5%/3mm DTA and 7%/4mm DTA were used in the study. The 3 irradiations for each insert were averaged together for comparison. RESULTS Film measurements for the 2D gamma showed 85% pixels passing at 3%/3mm in both planes. The 5%/3 mm constraint had 93% and 90% passing in the two planes. The 7%/4mm restraint resulted in 99% passing in both planes. The PRESAGE 2D gamma passed 66% and 61% of pixels in the both planes at 3%/3mm. At 5%/3mm 86% and 82% passed. For 7%/4mm, 94% of pixels passed in both planes. The 3D gamma resulted in a pass rate of 90% at 3%/3mm, 95% at 5%/3mm, and 99% at 7%/4mm. CONCLUSION 2D gamma pass rates using film showed a higher pass rate than PRESAGE using the same criteria in the same planes. This may be due to poor 3D registration with the treatment plan compared to the 2D film registration system. The 3D gamma results had a higher pass rate (> 90% pass rate) possibly because it sampled many more pixels in noncritical volumes thus diluting the percent of pixels passing. 3D restraints should be more restrictive to be comparable to 2D results. Funding from NIH grant 5R01CA100835.
منابع مشابه
Comparison of 2D and 3D Gamma calculations for an IMRT QA phantom
Gamma index pass rates were evaluated for an anthropomorphic phantom using both 2D and 3D calculations. The phantom was irradiated with the traditional dosimetry insert loaded with radiochromic film and TLD, and then with a 3D dosimetry insert. A comparison with the calculated dose distribution showed that both the PRESAGE ® dosimeter and the film and TLD system agreed with the plan to within 5...
متن کاملQuality Assurance / Phantoms
Gamma index pass rates were evaluated for an anthropomorphic phantom using both 2D and 3D calculations. The phantom was irradiated with the traditional dosimetry insert loaded with radiochromic film and TLD, and then with a 3D dosimetry insert. A comparison with the calculated dose distribution showed that both the PRESAGE® dosimeter and the film and TLD system agreed with the plan to within 5%...
متن کاملInvestigating the Feasibility of 3D Dosimetry in the RPC IMRT H&N Phantom.
An urgent requirement for 3D dosimetry has been recognized because of high failure rate (~25%) in RPC credentialing, which relies on point and 2D dose measurements. Comprehensive 3D dosimetry is likely to resolve more errors and improve IMRT quality assurance. This work presents an investigation of the feasibility of PRESAGE/optical-CT 3D dosimetry in the Radiologic Physics Center (RPC) IMRT H&...
متن کاملSU-E-T-73: Investigation of 3D Dosimetry for Proton Therapy Using PRESAGE.
PURPOSE This investigation studies the feasibility of PRESAGE, a 3D polyurethane dosimeter, for relative dosimetry measurements of clinically relevant proton treatments using an anthropomorphic head phantom developed by the Radiological Physics Center (RPC). Performance of a low-LET dependent PRESAGE was evaluated by comparison to the traditionally used radiochromic film, EBT2, and thermolumine...
متن کاملInvestigation of the feasibility of relative 3D dosimetry in the Radiologic Physics Center Head and Neck IMRT phantom using presage/optical-CT.
This study presents the application of the Presage/optical-CT 3D dosimetry system for relative dosimetry in the Radiologic Physics Center (RPC) Head and Neck (H&N) IMRT phantom. Performance of the system was evaluated by comparison with the "gold-standard" RPC credentialing test. A modified Presage cylindrical insert was created that extended the capability of the RPC H&N phantom to 3D dosimetr...