Brain perfusion CT compared with15O-H2O-PET in healthy subjects
نویسندگان
چکیده
BACKGROUND Regional cerebral blood flow [rCBF] measurements are valuable for identifying angiogenically active tumours, and perfusion computed tomography [CT] has been suggested for that purpose. This study aimed to validate rCBF measurements by perfusion CT with positron-emission tomography [PET] and15O-labelled water [15O-H2O] in healthy subjects. METHODS RCBF was measured twice in 12 healthy subjects with15O-H2O PET and once with perfusion CT performed over the basal ganglia. Matching rCBF values in regions of interest were compared. RESULTS Measured with perfusion CT, rCBF was significantly and systematically overestimated. White matter rCBF values were 17.4 ± 2.0 (mean ± SD) mL min-1 100 g-1 for PET and 21.8 ± 3.4 mL min-1 100 g-1 for perfusion CT. Grey matter rCBF values were 48.7 ± 5.0 mL min-1 100 g-1 for PET and 71.8 ± 8.0 mL min-1 100 g-1 for perfusion CT. The overestimation of grey matter rCBF could be reduced from 47% to 20% after normalization to white matter rCBF, but the difference was still significant. CONCLUSION RCBF measured with perfusion CT does contain perfusion information, but neither quantitative nor relative values can substitute rCBF measured by15O-H2O PET yet. This, however, does not necessarily preclude a useful role in patient management.
منابع مشابه
A statistical investigation of normal regional intra-subject heterogeneity of brain metabolism and perfusion by F-18 FDG and O-15 H2O PET imaging
BACKGROUND The definite evaluation of the regional cerebral heterogeneity using perfusion and metabolism by a single modality of PET imaging has not been well addressed. Thus a statistical analysis of voxel variables from identical brain regions on metabolic and perfusion PET images was carried out to determine characteristics of the regional heterogeneity of F-18 FDG and O-15 H2O cerebral upta...
متن کاملA Computed Tomography-Based Spatial Normalization for the Analysis of [18F] Fluorodeoxyglucose Positron Emission Tomography of the Brain
OBJECTIVE We developed a new computed tomography (CT)-based spatial normalization method and CT template to demonstrate its usefulness in spatial normalization of positron emission tomography (PET) images with [(18)F] fluorodeoxyglucose (FDG) PET studies in healthy controls. MATERIALS AND METHODS Seventy healthy controls underwent brain CT scan (120 KeV, 180 mAs, and 3 mm of thickness) and [(...
متن کاملQuantitative cerebral blood flow measurement with dynamic perfusion CT using the vascular-pixel elimination method: comparison with H2(15)O positron emission tomography.
BACKGROUND AND PURPOSE Blood vessels are usually conspicuous on dynamic CT perfusion images. The presence of large vessels may lead to overestimation of the quantitative value of cerebral blood flow (CBF). We evaluated the efficacy of the vascular-pixel elimination (VPE) method in quantitative CT perfusion imaging, in comparison with positron emission tomography (PET). METHODS Five healthy vo...
متن کاملEvaluation of role of brain SPECT in diagnosis of post stroke dementia [Persian]
Post stroke dementia is one of the most common complications of stroke that is preventable and relatively treatable too. The purpose of the study is comparison between the positive findings in brain CT scan and brain perfusion SPECT. 15 patients who were complicated by dementia after cerebrovascular accident and also 5 patients as a control group enrolled in this study. Brain CT scan and ...
متن کاملReference Ranges for LVEF and LV Volumes
Electrocardiographic gating is increasingly used for 82Rb cardiac PET/CT, but reference ranges for global functional parameters are not well defined. We sought to establish reference values for left ventricular ejection fraction (LVEF), end systolic volume (ESV), and end diastolic volume (EDV) using 4 different commercial software packages. Additionally, we compared 2 different approaches for t...
متن کامل