Central and Midperipheral Corneal Thickness Measured with Scheimpflug Imaging and Optical Coherence Tomography

نویسندگان

  • Jinhai Huang
  • Xixia Ding
  • Giacomo Savini
  • Zhengxuan Jiang
  • Chao Pan
  • Yanjun Hua
  • Fan Wu
  • Yifan Feng
  • Ye Yu
  • Qinmei Wang
چکیده

PURPOSE To compare corneal thickness measurements using Pentacam (Oculus, Germany), Sirius (CSO, Italy), Galilei (Ziemer, Switzerland), and RTVue-100 OCT (Optovue Inc., USA). METHODS Sixty-six eyes of 66 healthy volunteers were enrolled. Three consecutive measurements were performed with each device. The mean value of the three measurements was used for subsequent analysis. Central corneal thickness (CCT), thinnest corneal thickness (TCT), and midperipheral corneal thickness (MPCT; measured at superior, inferior, nasal, and temporal locations with a distance of 1 mm (CT2mm) or 2.5 mm (CT5mm) from the corneal apex) were analyzed. Differences and agreement between measurements were assessed using the repeated-measures analysis of variance (ANOVA) and Bland-Altman analyses, respectively. RESULTS Statistically significant differences (p<0.001) among the four devices were revealed in CCT, TCT and CT2mm measurements. The CCT, TCT, and CT2mm values were ranked from the thickest to the thinnest as follows: Galilei>Sirius>Pentacam>RTVue OCT. For these measurements, agreement between measurements by Sirius and Pentacam was good, whereas Galilei overestimated and RTVue underestimated corneal thickness compared to Sirius and Pentacam. As regards CT5mm measurements, Pentacam provided the largest values, whereas RTVue OCT yielded the smallest values. Agreement of the CT5mm measurements was good between the Pentacam, Sirius, moderate between Galilei and the other two Scheimpflug systems, and poor between the RTVue OCT and the remaining devices. CONCLUSIONS The Pentacam and Sirius can be used interchangeably for CCT measurements, while the Galilei and RTVue systematically over- and underestimate CCT, respectively. The three Scheimpflug cameras, but not the RTVue, may be used interchangeably for MPCT measurements.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Longitudinal Evaluation of Cornea With Swept-Source Optical Coherence Tomography and Scheimpflug Imaging Before and After Lasik

Swept-source optical coherence tomography (OCT) is the latest advancement in anterior segment imaging. There are limited data regarding its performance after laser in situ keratomileusis (LASIK). We compared the reliability of swept-source OCT and Scheimpflug imaging for evaluation of corneal parameters in refractive surgery candidates with myopia or myopic astigmatism. Three consecutive measur...

متن کامل

3D refraction correction and extraction of clinical parameters from spectral domain optical coherence tomography of the cornea.

Capable of three-dimensional imaging of the cornea with micrometer-scale resolution, spectral domain-optical coherence tomography (SDOCT) offers potential advantages over Placido ring and Scheimpflug photography based systems for accurate extraction of quantitative keratometric parameters. In this work, an SDOCT scanning protocol and motion correction algorithm were implemented to minimize the ...

متن کامل

Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology

PURPOSE The purpose of this study was to compare the central corneal thickness (CCT) measurements of four noncontact devices in healthy eyes. MATERIALS AND METHODS In a sample of 45 healthy controls, CCT was measured using an optical biometer (IOLMaster 700) based on swept-source optical coherence tomography (SS-OCT), high-resolution rotating Scheimpflug camera system (Pentacam HR), spectral-...

متن کامل

Vertical and horizontal thickness profiles of the corneal epithelium and Bowman's layer after orthokeratology.

PURPOSE To investigate thickness profile changes of the corneal epithelium and Bowman's layer at the vertical and horizontal meridians with overnight myopia orthokeratology (OK) lenses. METHODS Twenty subjects (age range: 19-33 years) wore reverse-geometry rigid gas-permeable OK lenses in both eyes for 30 days. Before lens wear and after 1, 7, and 30 days of overnight lens wear, evaluation of...

متن کامل

Age- and sex-related differences in corneal epithelial thickness measured with spectral domain anterior segment optical coherence tomography among Egyptians

Is to measure the corneal epithelial thickness (CET) in a group of healthy Egyptian population and to investigate its variation with age and sex using spectral domain-anterior segment optical coherence tomography (SD-AS-OCT).This cross-sectional observational study includes 240 eyes of 120 healthy individuals (60 men, 60 women). Each gender was divided into 4 equal groups according to age as fo...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 9  شماره 

صفحات  -

تاریخ انتشار 2014