Performance of Gem Premier blood gas/electrolyte analyzer evaluated.

نویسندگان

  • E Jacobs
  • M Nowakowski
  • N Colman
چکیده

We evaluated a new analyzer designed for point-of-care testing of blood gases, sodium, potassium, ionized calcium, and hematocrit. The Gem Premier (Mallinckrodt) system has two components: the analyzer and a disposable cartridge. Analysis takes place in the cartridge, which contains the electrochemical sensors, the calibrants, the reagents, the sampling stylus, and the waste container. The system was evaluated for imprecision and accuracy. With aqueous control materials, total imprecision (CV) was: pH, 0.10-0.18%; PCO2, 3.16-5.78%; PO2, 2.92-4.85%; sodium, 0.82-1.44%; potassium, 1.35-1.48%; ionized calcium, 0.75-1.45%; and hematocrit, 1.13-1.83%. Accuracy of the system was assessed by split-sample comparison with the Radiometer ABL 330 blood gas analyzer for pH and blood gases, the Nova Stat Profile 5 for whole-blood electrolyte and hematocrit analysis, and the IL Phoenix for plasma electrolyte analysis. After outlier correction, regression statistics were excellent for all analytes except sodium, which demonstrated Sy[x values between 1.80 and 2.30 mmol/L and 0.85 < or = r < 0.90.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of blood gas and electrolyte test results from the Gem-Premier and the ABL-70 versus a conventional laboratory analyzer.

Blood gas analyzers serve a critical role in providing information that reflects patient homeostasis. This study was undertaken to evaluate the accuracy, reliability, consistency, and bias of the Radiometer ABL-70 point of care blood gas analyzer. Thirty samples were gathered for analysis of pH, pCO2, pO2, sodium, potassium, sodium bicarbonate, and base excess. Twenty-nine samples were gathered...

متن کامل

Validation of a quality assessment system for blood gas and electrolyte testing.

BACKGROUND A recently-introduced quality assessment system (Intelligent Quality Management: iQM), was evaluated in routine clinical use at four different hospitals. The iQM technology is designed to replace conventional external liquid controls with software, Process Control (PC) Solutions and Calibration Validation components that continually assess the function of the GEM Premier 3000 (GEM) a...

متن کامل

Comparison of blood gas and electrolyte test results from the Gem-Stat and CDI-300 versus a conventional laboratory analyzer.

Continuous blood gas monitoring devices have been an aid to the perfusionist since the introduction of the oxygen saturation meters of the early 1980s. Since that time, the perfusionist has had to decide between continuous versus intermittent sampling, and on-line (an analyzer that can automatically sample either at prescribed intervals and/or on demand) versus in-line devices (monitors that c...

متن کامل

Limitations and opportunities of whole blood bilirubin measurements by GEM premier 4000®

BACKGROUND Neonatal hyperbilirubinemia has traditionally been screened by either total serum bilirubin or transcutaneous bilirubin. Whole blood bilirubin (TwB) by the GEM Premier 4000® blood gas analyzer (GEM) is a relatively new technology and it provides fast bilirubin results with a small sample volume and can measure co-oximetry and other analytes. Our clinical study was to evaluate the rel...

متن کامل

Method validation of a set of 12 GEM® Premier™ 4000 blood gas analyzers for point-of-care testing in a university teaching hospital

Background Blood gas analyzers are o0.ften integrated into point-of-care testing provisions. International standards (ISO 22870 and 15189) as adapted to French COFRAC regulations make accreditation of point-ofta-care testintag obligatory. We installed and assessed 12 GEM PREMIER 4000 analyzers for pH, pCO2, pO2, Na+, K+, Cl-, Ca2+, lactate, hemoglobin and oxyhemoglobin (O2Hb) at Clermont-Ferran...

متن کامل