Uncertainty Analysis in Humidity Measurements by the Psychrometer Method
نویسندگان
چکیده
The most common and cheap indirect technique to measure relative humidity is by using psychrometer based on a dry and a wet temperature sensor. In this study, the measurement uncertainty of relative humidity was evaluated by this indirect method with some empirical equations for calculating relative humidity. Among the six equations tested, the Penman equation had the best predictive ability for the dry bulb temperature range of 15-50 °C. At a fixed dry bulb temperature, an increase in the wet bulb depression increased the error. A new equation for the psychrometer constant was established by regression analysis. This equation can be computed by using a calculator. The average predictive error of relative humidity was <0.1% by this new equation. The measurement uncertainty of the relative humidity affected by the accuracy of dry and wet bulb temperature and the numeric values of measurement uncertainty were evaluated for various conditions. The uncertainty of wet bulb temperature was the main factor on the RH measurement uncertainty.
منابع مشابه
An Inexpensive, Stable, and Accurate Relative Humidity Measurement Method for Challenging Environments
In this research, an improved psychrometer is developed to solve practical issues arising in the relative humidity measurement of challenging drying environments for meat manufacturing in agricultural and agri-food industries. The design in this research focused on the structure of the improved psychrometer, signal conversion, and calculation methods. The experimental results showed the effect ...
متن کاملVentilation effects on humidity measurements in thermometer screens
Relative humidity (RH) measurements, as derived from wet-bulb and dry-bulb thermometers operated as a psychrometer within a thermometer screen, have limited accuracy because of natural ventilation variations. Standard RH calculations generally assume a fixed screen psychrometer coefficient, but this is too small during poor ventilation. By comparing a reference humidity probe—exposed within a s...
متن کاملEvaluation of Effective Sources in Uncertainty Measurements of Personal Dosimetry by a Harshaw TLD System
Background: The accurate results of the individual doses in personal dosimety which are reported by the service providers in personal dosimetry are very important. There are national / international criteria for acceptable dosimetry system performance.Objective: In this research, the sources of uncertainties are identified, measured and calculated in a personal dosimetry system by TLD.Method: T...
متن کاملA Differential Psychrometer for Continuous Measurements of Transpiration.
Transpiration is frequently measured by enclosing plants, or portions of plants, in transparent chambers and determining the change of humidity of a stream of air before and after passing through the enclosure (2, 4). Although this procedure should not be used to estimate probable transpiration rates under natural conditions, unless adequate precautions are taken to ensure that the microenviron...
متن کاملComparison of the dye method with the thermocouple psychrometer for measuring leaf water potentials.
The dye method for measuring water potential was examined and compared with the thermocouple psychrometer method in order to evaluate its usefulness for measuring leaf water potentials of forest trees and common laboratory plants. Psychrometer measurements are assumed to represent the true leaf water potentials. Because of the contamination of test solutions by cell sap and leaf surface residue...
متن کامل