An Analytical Approach towards Passive Ferromagnetic Shimming Design for a High-Resolution NMR Magnet.

نویسندگان

  • Frank X Li
  • John P Voccio
  • Min Cheol Ahn
  • Seungyong Hahn
  • Juan Bascuñán
  • Yukikazu Iwasa
چکیده

This paper presents a warm bore ferromagnetic shimming design for a high resolution NMR magnet based on spherical harmonic coefficient reduction techniques. The passive ferromagnetic shimming along with the active shimming is a critically important step to improve magnetic field homogeneity for an NMR Magnet. Here, the technique is applied to an NMR magnet already designed and built at the MIT's Francis Bitter Magnet Lab. Based on the actual magnetic field measurement data, a total of twenty-two low order spherical harmonic coefficients is derived. Another set of spherical harmonic coefficients was calculated for iron pieces attached to a 54 mm diameter and 72 mm high tube. To improve the homogeneity of the magnet, a multiple objective linear programming method was applied to minimize unwanted spherical harmonic coefficients. A ferromagnetic shimming set with seventy-four iron pieces was presented. Analytical comparisons are made for the expected magnetic field after Ferromagnetic shimming. The theoretically reconstructed magnetic field plot after ferromagnetic shimming has shown that the magnetic field homogeneity was significantly improved.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

High-field NMR using resistive and hybrid magnets.

Resistive and resistive-superconducting hybrid magnets can generate dc magnetic fields much higher than conventional superconducting NMR magnets but the field spatial homogeneity and temporal stability are usually not sufficient for high-resolution NMR experiments. Hardware and technique development addressing these issues are presented for high-resolution NMR at magnetic fields up to 40T. Pass...

متن کامل

Shimming Permanent Magnet of MRI Scanner

In this paper an approach of ferromagnetic shimming for permanent MRI magnet is presented. It is designed to reduce unhomogeneity of magnetostatic field of permanent magnet to meet the stringent requirement for MRI applications. An optimal configuration of ferromagnetic pieces is generated through calculation according to the initial field map and the demanded final homogeneity specifications. ...

متن کامل

Paramagnetic shimming for wide-range variable-field NMR.

We propose a new passive shimming strategy for variable-field NMR experiments, in which the magnetic field produced by paramagnetic shim pieces placed inside the magnet bore compensates the inhomogeneity of a variable-field magnet for a wide range of magnet currents. Paramagnetic shimming is demonstrated in (7)Li, (87)Rb, and (45)Sc NMR of a liquid solution sample in magnetic fields of 3.4, 4.0...

متن کامل

The Shimming of High Resolution NMR Magnets

This is the companion to the shimming simulation module contained in the NUTS NMR data processing software.

متن کامل

Multipole shimming of permanent magnets using harmonic corrector rings.

Shimming systems are required to provide sufficient field homogeneity for high resolution nuclear magnetic resonance (NMR). In certain specialized applications, such as rotating-field NMR and mobile ex situ NMR, permanent magnet-based shimming systems can provide considerable advantages. We present a simple two-dimensional shimming method based on harmonic corrector rings which can provide arbi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Superconductor science & technology

دوره 28 7  شماره 

صفحات  -

تاریخ انتشار 2015