Finite element analysis of cementless femoral stems based on mid- and long-term radiological evaluation

نویسندگان

  • Kanehiro Matsuyama
  • Yasuhiro Ishidou
  • Yong-Ming Guo
  • Hironori Kakoi
  • Takao Setoguchi
  • Satoshi Nagano
  • Ichiro Kawamura
  • Shingo Maeda
  • Setsuro Komiya
چکیده

BACKGROUND Femoral bone remodeling in response to stress shielding induces periprosthetic bone loss. Computerized finite element analysis (FEA) is employed to demonstrate differences in initial stress distribution. However, FEA is often performed without considering the precise sites at which the stem was fixed. We determined whether FEA reflects mid-term radiological examination exactly as predicted following long-term stress shielding. METHODS Femur-stem fixation sites were evaluated radiologically according to the location of spot welds in two anatomical cementless stem designs. Based on mid-term radiological results, four femur-stem bonding site conditions were defined as: (Condition A) no bonding; (Condition B) bonding within the 10 mm area proximal to the distal border of the porous area; (Condition C) bonding of the entire porous area; and (Condition D) bonding of the entire femoral stem, prior to conducting FEA analysis. Furthermore, we radiographically evaluated mid- and long-term stress shielding, and measured bone mineral density of the femur 10 years after total hip arthroplasty. RESULTS Spot welds appeared frequently around the border between the porous and smooth areas. FEA showed that, based on mid-term radiological evaluation, von Mises stress was reduced in condition B in the area proximal to the femur-stem bonding sites for both stem designs compared with condition A (no bonding). Conversely, von Mises stress at all areas of the femur-stem bonding sites in conditions C and D was higher than that in condition A. With respect to stress shielding progression, there was no significant difference between the two types of stem designs. However, stress shielding progressed and was significantly higher in the presence of spot welds (p = 0.001). In both stem designs, bone mineral density in zone VII was significantly lower than that in the contralateral hips. CONCLUSIONS These results indicate that FEA based on mid-term radiological evaluation may be helpful to predict the influence of long-term stress shielding more precisely.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The Effect of Physiological Load Configuration on Interface Micromotion in Cementless Femoral Stems

The most commonly reported failure modes of cementless hip stems are loosening and thigh pain; both are attributed to the relative motion at the bone-implant interface due to failure to achieve sufficient primary fixation. Accurate predictions of hip stems’ stability are therefore crucial to the pre-clinical analyses of hip arthroplasty. This study uses finite element technique to analyse the e...

متن کامل

Mid-term outcome of a modular, cementless, proximally hydroxyapatite-coated, anatomic femoral stem.

PURPOSE To report the mid-term outcome of a modular, cementless, proximally hydroxyapatitecoated, anatomic femoral stem in total hip arthroplasty (THA). METHODS 160 consecutive patients aged 42 to 92 (mean, 70) years underwent 185 cementless THAs for primary osteoarthritis or femoral neck fractures. All procedures were performed by a single surgeon using the same modular, cementless, proximal...

متن کامل

A Finite Element Analysis of the Effects of Bone Loss on Cementless Total Hip Arthroplasty

Fatigue fractures of the femoral component of cementless hip prostheses are rare occurrences at present, but may become more widespread in the future as the demand increases for porous coated cementless hip arthroplasty in young, active, and heavy patients. Whereas the fatigue fracture of high strength forged alloys is unlikely, a risk of fatigue fracture may exist for sintered porous coated st...

متن کامل

The modern, hybrid total hip arthroplasty for primary osteoarthritis at the Hospital for Special Surgery.

We describe our technique and rationale using hybrid fixation for primary total hip arthroplasty (THA) at the Hospital for Special Surgery. Modern uncemented acetabular components have few screw holes, or no holes, polished inner surfaces, improved locking mechanisms, and maximised thickness and shell-liner conformity. Uncemented sockets can be combined with highly cross-linked polyethylene lin...

متن کامل

Primary Stability Recognition of the Newly Designed Cementless Femoral Stem Using Digital Signal Processing

Stress shielding and micromotion are two major issues which determine the success of newly designed cementless femoral stems. The correlation of experimental validation with finite element analysis (FEA) is commonly used to evaluate the stress distribution and fixation stability of the stem within the femoral canal. This paper focused on the applications of feature extraction and pattern recogn...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 17  شماره 

صفحات  -

تاریخ انتشار 2016