Study of scattered photons from the collimator system of Leksell Gamma Knife using the EGS4 Monte Carlo code.
نویسندگان
چکیده
In the algorithm of Leksell GAMMAPLAN (the treatment planning software of Leksell Gamma Knife), scattered photons from the collimator system are presumed to have negligible effects on the Gamma Knife dosimetry. In this study, we used the EGS4 Monte Carlo (MC) technique to study the scattered photons coming out of the single beam channel of Leksell Gamma Knife. The PRESTA (Parameter Reduced Electron-Step Transport Algorithm) version of the EGS4 (Electron Gamma Shower version 4) MC computer code was employed. We simulated the single beam channel of Leksell Gamma Knife with the full geometry. Primary photons were sampled from within the 60Co source and radiated isotropically in a solid angle of 4pi. The percentages of scattered photons within all photons reaching the phantom space using different collimators were calculated with an average value of 15%. However, this significant amount of scattered photons contributes negligible effects to single beam dose profiles for different collimators. Output spectra were calculated for the four different collimators. To increase the efficiency of simulation by decreasing the semiaperture angle of the beam channel or the solid angle of the initial directions of primary photons will underestimate the scattered component of the photon fluence. The generated backscattered photons from within the 60Co source and the beam channel also contribute to the output spectra.
منابع مشابه
Monte Carlo calculations and GafChromic film measurements for plugged collimator helmets of Leksell Gamma Knife unit.
The Monte Carlo technique and GafChromic films were employed to verify the accuracy of the dose planning system (Leksell GammaPlan) used in Gamma Knife (type B) radiosurgery when plugged collimator helmets were used. The EGS4 Monte Carlo code was used to calculate the dose distribution along the x, y, and z axes when a single shot was delivered at the center point (unit center point: x = 100, y...
متن کاملDose distributions at extreme irradiation depths of gamma knife radiosurgery: EGS4 Monte Carlo calculations.
The accuracy of the dose planning system (Leksell GammaPlan), used in Gamma Knife (type B) radiosurgery at extreme irradiation depths, was verified using the Monte Carlo technique. EGS4 Monte Carlo calculations were employed to calculate the dose distribution along the x, y and z axes for an irradiation relatively shallow in a spherical bony cavity water phantom. Two different sizes of the coll...
متن کاملMonte Carlo Evaluation of Gamma Knife Dose Profile in Real Brain Phantom
Introduction The Gamma Knife system is designed solely for non-invasive treatment of brain disorders, and it benefits from stereotactic surgical techniques. Dose calculations required in the system are performed by GammaPlan code; in this code, brain tissue is considered uniform. In the present study, we evaluated the effect of Gamma Knife system on the obtained dose through simulating a real ...
متن کاملMonte Carlo calculation of single-beam dose profiles used in a gamma knife treatment planning system.
The accuracy of single-beam dose profiles used in the algorithm of the Gamma Knife treatment planning system (Leksell GammaPlan) is verified. EGS4 Monte Carlo calculation was employed to calculate the dose distributions of single-beams in a spherical water phantom with diameter 160 mm. The beams were directed to the center of the phantom. Collimators of 4, 8, 14, and 18 mm sizes were studied. T...
متن کاملMonte Carlo calculated output factors of a Leksell Gamma Knife unit.
The Leksell Gamma Knife is a standard radiosurgical tool for treating brain lesions by directing beams of gamma radiation to a specific region. The diameter of the gamma beams is confined by collimator systems and available collimator sizes are 4, 8, 14 and 18 mm. The reduction in dose rate for each collimator helmet is called the output factor (OPF). Experimental determination of OPFs is diffi...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Medical physics
دوره 33 1 شماره
صفحات -
تاریخ انتشار 2006