An Indirect Method to Measure Abutment Screw Preload: A Pilot Study Based on Micro-CT Scanning.
نویسندگان
چکیده
This study aimed to measure the preload in different implant platform geometries based on micro-CT images. External hexagon (EH) implants and Morse Tapered (MT) implants (n=5) were used for the preload measurement. The abutment screws were scanned in micro-CT to obtain their virtual models, which were used to record their initial length. The abutments were screwed on the implant with a 20 Ncm torque and the set composed by implant, abutment screw and abutment were taken to the micro-CT scanner to obtain virtual slices of the specimens. These slices allowed the measurement of screw lengths after torque application and based on the screw elongation. Preload values were calculated using the Hooke's Law. The preloads of both groups were compared by independent t-test. Removal torque of each specimen was recorded. To evaluate the accuracy of the micro-CT technique, three rods with known lengths were scanned and the length of their virtual model was measured and compared with the original length. One rod was scanned four times to evaluate the measuring method variation. There was no difference between groups for preload (EH = 461.6 N and MT = 477.4 N), but the EH group showed higher removal torque values (13.8 ± 4.7 against 8.2 ± 3.6 N cm for MT group). The micro-CT technique showed a variability of 0.053% and repeatability showed an error of 0.23 to 0.28%. Within the limitations of this study, there was no difference between external hexagon and Morse taper for preload. The method using micro-CT may be considered for preload calculation.
منابع مشابه
Preload loss in gold prosthesis-retaining screws as a function of time.
PURPOSE Screwed joints are widely used in implant dentistry, and their failure is a significant problem that may be related to loss of preload. Preload is the compressive force generated across a joint when a screw is tightened and is responsible for keeping a joint closed. For a given torque, preload is limited by the frictional resistance of the contacting screw threads, flange, and opposing ...
متن کاملImplant-abutment screw joint preload of 7 hex-top abutment systems.
This study measured the screw joint preload of the implant-abutment interface for 7 common hex-top abutment systems. Increasingly, prosthetic designs are utilizing a direct prosthetic connection to the implant, placing the implant-abutment screw joint under direct functional loads and moments. Sufficiently high screw joint preloads are required to maintain screw joint integrity and confer clini...
متن کاملMeasuring abutment/implant joint integrity with the Periotest instrument.
Maintenance of the integrity of the abutment/implant interface is essential and is dependent on the abutment screw retaining a preload. Evaluation of this joint is usually done by manual assessment. The purpose of the current study was to determine whether the Periotest instrument could be used to evaluate abutment screw loosening. A custom-designed apparatus was constructed to measure abutment...
متن کاملFixture-abutment connection surface and micro-gap measurements by 3D micro-tomographic technique analysis.
X-ray micro-tomography (micro-CT) is a miniaturized form of conventional computed axial tomography (CAT) able to investigate small radio-opaque objects at a-few-microns high resolution, in a non-destructive, non-invasive, and tri-dimensional way. Compared to traditional optical and electron microscopy techniques, which provide two-dimensional images, this innovative investigation technology ena...
متن کاملInfluence of lubricant on screw preload and stresses in a finite element model for a dental implant.
STATEMENT OF PROBLEM Loosening or fracture of the abutment screw are frequent complications in implant dentistry and are detrimental to the long-term success of the restorations. However, little is known about the factors influencing the stability of the screw-abutment complex. PURPOSE The purpose of this study was to investigate the influence of lubricant action during implant assembly on sc...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Brazilian dental journal
دوره 26 6 شماره
صفحات -
تاریخ انتشار 2015