Multi-focal HIFU reduces cavitation in mild-hyperthermia
نویسندگان
چکیده
BACKGROUND Mild-hyperthermia therapy (40-45 °C) with high-intensity focused ultrasound (HIFU) is a technique being considered in a number of different treatments such as thermally activated drug delivery, immune-stimulation, and as a chemotherapy adjuvant. Mechanical damage and loss of cell viability associated with HIFU-induced acoustic cavitation may pose a risk during these treatments or may hinder their success. Here we present a method that achieves mild heating and reduces cavitation by using a multi-focused HIFU beam. We quantify cavitation level and temperature rise in multi-focal sonications and compare it to single-focus sonications at the transducer geometric focus. METHODS Continuous wave sonications were performed with the Sonalleve V2 transducer in gel phantoms and pork at 5, 10, 20, 40, 60, 80 acoustic watts for 30 s. Cavitation activity was measured with two ultrasound (US) imaging probes, both by computing the raw channel variance and using passive acoustic mapping (PAM). Temperature rise was measured with MR thermometry at 3 T. Cavitation and heating were compared for single- and multi-focal sonication geometries. Multi-focal sonications used four points equally spaced on a ring of either 4 mm or 8 mm diameter. Single-focus sonications were not steered. RESULTS Multi-focal sonication generated distinct foci that were visible in MRI thermal maps in both phantoms and pork, and visible in PAM images in phantoms only. Cavitation activity (measured by channel variance) and mean PAM image value were highly correlated (r > 0.9). In phantoms, cavitation exponentially decreased over the 30-second sonication, consistent with depletion of cavitation nuclei. In pork, sporadic spikes signaling cavitation were observed with single focusing only. In both materials, the widest beam reduced average and peak cavitation level by a factor of two or more at each power tested when compared to a single focus. The widest beam reduced peak temperature by at least 10 °C at powers above 5 W, and created heating that was more spatially diffuse than single focus, resulting in more voxels in the mild heating (3-8 °C) range. CONCLUSIONS Multi-focal HIFU can be used to achieve mild temperature elevation and reduce cavitation activity.
منابع مشابه
Enhanced cavitation by using two consecutive ultrasound waves at different frequencies.
Efficient and noninvasive generation of cavitation bubbles in soft tissue is a challenging task due to the lack of cavitation nuclei (i.e., pre-existing gas bubbles). In this study, we present a method to generate and enhance cavitation activity based on the utilization of two consecutive ultrasound waves at different frequencies. First, a high frequency (5 MHz) high intensity focused ultrasoun...
متن کاملDetection and exploitation of acoustic cavitation for enhancement of MR guided High Intensity Focused Ultrasound heating in ex vivo liver
Purpose/Introduction High Intensity Focused Ultrasound (HIFU) is increasingly used for the non invasive treatment of pathological tissue. The efficiency of heating can be increased by generation of acoustic cavitation in order to modify the local distribution of ultrasound pressure that can be visualized with Magnetic Resonance Acoustic Force Imaging (MR-ARFI). This studies analyzes the change ...
متن کاملEffects of HIFU induced cavitation on flooded lung parenchyma
BACKGROUND High intensity focused ultrasound (HIFU) has gained clinical interest as a non-invasive local tumour therapy in many organs. In addition, it has been shown that lung cancer can be targeted by HIFU using One-Lung Flooding (OLF). OLF generates a gas free saline-lung compound in one lung wing and therefore acoustic access to central lung tumours. It can be assumed that lung parenchyma i...
متن کاملEnhancement of High-Intensity Focused Ultrasound Heating by Short-Pulse Generated Cavitation
A target tissue can be thermally coagulated in high-intensity focused ultrasound (HIFU) treatment noninvasively. HIFU thermal treatments have been clinically applied to various solid tumors. One of the problems in HIFU treatments is a long treatment time. Acoustically driven microbubbles can accelerate the ultrasonic heating, resulting in the significant reduction of the treatment time. In this...
متن کاملUltrasound Exposimetry: Are Current Clinical Instruments Capable of Causing Harm to the Human Embryo or Fetus?
Transrectal High Intensity Focused Ultrasound treatment is currently being studied as a minimally invasive therapy of localized prostatic cancer. In is a novel therapeutic procedure in which a highly focused transducer induces tissue coagulation necrosis by sharp temperature increase inside the prostatic tissue. Generally accepted belief holds that the tissue necrosis is caused by two principal...
متن کامل