The creep and wear of highly cross-linked polyethylene: a three-year randomised, controlled trial using radiostereometric analysis.

نویسندگان

  • S Glyn-Jones
  • P McLardy-Smith
  • H S Gill
  • D W Murray
چکیده

The creep and wear behaviour of highly cross-linked polyethylene and standard polyethylene liners were examined in a prospective, double-blind randomised, controlled trial using radiostereometric analysis. We randomised 54 patients to receive hip replacements with either highly cross-linked polyethylene or standard liners and determined the three-dimensional penetration of the liners over three years. After three years the mean total penetration was 0.35 mm (SD 0.14) for the highly cross-linked polyethylene group and 0.45 mm (SD 0.19) for the standard group. The difference was statistically significant (p = 0.0184). From the pattern of penetration it was possible to discriminate creep from wear. Most (95%) of the creep occurred within six months of implantation and nearly all within the first year. There was no difference in the mean degree of creep between the two types of polyethylene (highly cross-linked polyethylene 0.26 mm, SD 0.17; standard 0.27 mm, SD 0.2; p = 0.83). There was, however, a significant difference (p = 0.012) in the mean wear rate (highly cross-linked polyethylene 0.03 mm/yr, SD 0.06; standard 0.07 mm/yr, SD 0.05). Creep and wear occurred in significantly different directions (p = 0.01); creep was predominantly proximal whereas wear was anterior, proximal and medial. We conclude that penetration in the first six months is creep-dominated, but after one year virtually all penetration is due to wear. Highly cross-linked polyethylene has a 60% lower rate of wear than standard polyethylene and therefore will probably perform better in the long term.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Cross-linked polyethylene and bisphosphonate therapy for osteolysis in total hip arthroplasty: a case report.

A 39-year-old woman underwent bilateral total hip arthroplasty with conventional, ethylene oxide-sterilised liners when she was a subject in a radiostereometric analysis study. Within 2 years she had rapid polyethylene wear with aggressive, asymptomatic, and periprosthetic osteolysis on both sides. Oral alendronate therapy halted the progression of osteolysis over a year and revision to cross-l...

متن کامل

Five-Year Experience of Vitamin E-Diffused Highly Cross-Linked Polyethylene Wear in Total Hip Arthroplasty Assessed by Radiostereometric Analysis.

BACKGROUND Vitamin E-diffused highly cross-linked polyethylene (VEPE) was developed to reduce oxidation without compromising mechanical strength. The purpose of this study was to evaluate VEPE in vivo using radiostereometric analysis (RSA) and patient-reported outcome measures (PROMs). METHODS Fifty-one hips were enrolled. Each patient received a VEPE liner, a porous titanium shell, and an un...

متن کامل

A prospective randomised controlled trial comparing three alternative bearing surfaces in primary total hip replacement.

The ideal bearing surface for young patients undergoing total hip replacement (THR) remains controversial. We report the five-year results of a randomised controlled trial comparing the clinical and radiological outcomes of 102 THRs in 91 patients who were < 65 years of age. These patients were randomised to receive a cobalt-chrome on ultra-high-molecular-weight polyethylene, cobalt-chrome on h...

متن کامل

Five-year comparison of wear using oxidised zirconium and cobalt-chrome femoral heads in total hip arthroplasty: a multicentre randomised controlled trial.

Oxidised zirconium (OxZi) has been developed as an alternative bearing surface for femoral heads in total hip arthroplasty (THA). This study has investigated polyethylene wear, functional outcomes and complications, comparing OxZi and cobalt-chrome (CoCr) as part of a three-arm, multicentre randomised controlled trial. Patients undergoing THA from four institutions were prospectively randomised...

متن کامل

Radiostereometric analysis study of tantalum compared with titanium acetabular cups and highly cross-linked compared with conventional liners in young patients undergoing total hip replacement.

BACKGROUND Radiostereometric analysis provides highly precise measurements of component micromotion relative to the bone that is otherwise undetectable by routine radiographs. This study compared, at a minimum of five years following surgery, the micromotion of tantalum and titanium acetabular cups and femoral head penetration in highly cross-linked polyethylene liners and conventional (ultra-h...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The Journal of bone and joint surgery. British volume

دوره 90 5  شماره 

صفحات  -

تاریخ انتشار 2008