Off-pump transapical mitral valve-in-ring implantation.

نویسندگان

  • Yu Zou
  • Enrico Ferrari
  • Ludwig K von Segesser
چکیده

OBJECTIVES The study aimed to evaluate the feasibility of off-pump transapical mitral valve-in-ring implantation and to test the performance of a custom-made self-expandable stent valve, in comparison with the standard SAPIEN valve. METHODS Acute experiments were performed in five pigs. Animals (mean weight 58.4 ± 7.3 kg) underwent mitral valve annuloplasties under cardiopulmonary bypass using 26-mm rings (SJM™). Then, a 30-mm custom-made self-expandable stent valve or a 23-mm balloon-expandable transcatheter heart valve (Edwards SAPIEN XT™) was deployed within the annuloplasty rings through a transatrial access and under direct vision. Subsequently, the stent valves were inserted transapically under fluoroscopic guidance and off pump. RESULTS The procedural success of transatrial and transapical mitral valve-in-ring procedures was 100% (10 of 10). Mean transatrial and transapical procedure time was 2.0 ± 1.1 and 22.0 ± 5.7 min, respectively. Haemodynamic status during transapical implantation remained stable, and differences in data collected before and after the stent-valve deployment were not statistically significant. Mean mitral annulus diameter and mean mitral orifice area in the group of self-expandable stent valves were 2.60 ± 0.02 cm and 4.16 ± 0.48 cm(2), respectively, whereas in the SAPIEN group they were 1.95 ± 0.18 cm and 2.26 ± 0.20 cm(2), respectively. Trace or mild regurgitation was detected only in the self-expandable stent-valve group. Mean gradients were 4.1 ± 4.5 mmHg across the self-expandable stent valves and 1.0 ± 0 mmHg across the SAPIEN valves. Postmortem examination confirmed adequate positioning of the self-expandable valves and the SAPIEN valves within the annuloplasty ring. CONCLUSIONS Off-pump transapical mitral valve-in-ring implantation is safe and feasible. Transapical access may represent the ideal option for valve-in-ring procedures in cases of recurrent mitral regurgitation after mitral valve repair, in high-risk patients. Owing to the supra-annular profile of the valve components, our custom-made nitinol stent valve provides nearer to normal functional area than the SAPIEN valve.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Off-pump transapical implantation of artificial neo-chordae to correct mitral regurgitation: the TACT Trial (Transapical Artificial Chordae Tendinae) proof of concept.

OBJECTIVES The goal of this study was to evaluate the safety and performance of the NeoChord DS1000 system (NeoChord, Inc., Minneapolis, Minnesota). BACKGROUND There is an increasing interest in transcatheter mitral valve (MV) treatment. The NeoChord DS 1000 system enables off-pump beating heart transapical MV repair with implantation of artificial neo-chordae. METHODS Patients with severe ...

متن کامل

New Self-Expanding Transcatheter Valve for Off-Pump Transatrial Mitral Valve-In-Ring Implantation.

OBJECTIVES To validate a self-expanding transcatheter valve for off-pump transatrial mitral valve-in-ring (VIR) implantation via a left thoracotomy. METHODS Mitral valve annuloplasty was performed via sternotomy during cardiopulmonary bypass on 9 pigs. After successful weaning from extracorporal circulation, the custom-made, self-expanding transcatheter VIR device was deployed under fluorosco...

متن کامل

Minimally invasive off-pump valve-in-a-ring implantation: the atrial transcatheter approach for re-operative mitral valve replacement after failed repair.

OBJECTIVE Based upon recent developments in transcatheter technology, this study was designed to evaluate the feasibility and haemodynamic performance of transcatheter valve-in-a-ring (VinR) implantation for potentially failed mitral repair using a minimally invasive, transatrial, off-pump approach. METHODS Adult sheep (54.3+/-3.0 kg) underwent mitral valve repair with a 26 mm complete annulo...

متن کامل

Feasibility of anterior mitral leaflet flail repair with transapical beating-heart neochord implantation.

A 91-year-old man was referred following congestive heart failure with acute pulmonary edema that was medically treated. Echocardiography revealed the presence of severe mitral regurgitation (MR) as a result of chordal rupture on the anterior leaflet at the level of the A2 segment (Figures 1A to 1C, Online Videos 1, 2 and 3). He presented with a history of arterial hypertension, hypothyroidism,...

متن کامل

Transapical mitral valve implantation: the Lutter valve

  The development of transcatheter techniques for treatment of severe mitral valve regurgitation in the beating heart is focus of recent research. An off-pump treatment technique poses great benefits, particularly for multimorbid patients, often being non-compliant to the gold standard treatment, being open heart surgery with use of a cardiopulmonary bypass. Thereto, two approaches are being fo...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery

دوره 43 4  شماره 

صفحات  -

تاریخ انتشار 2013