Long-term outcome of a cementless, hemispherical, press-fit acetabular component: survivorship analysis and dose-response relationship to linear polyethylene wear.

نویسندگان

  • N W Emms
  • I Stockley
  • A J Hamer
  • J M Wilkinson
چکیده

Between 1988 and 1998 we implanted 318 total hip replacements (THRs) in 287 patients using the Plasmacup (B. Braun Ltd, Sheffield, United Kingdom) and a conventional metal-on-polyethylene articulation. The main indications for THR were primary or secondary osteoarthritis. At follow-up after a mean 11.6 years (7.6 to 18.4) 17 patients had died and 20 could not be traced leaving a final series of 280 THRs in 250 patients. There were 62 revisions (22.1%) in 59 patients. A total of 43 acetabular shells (15.4%) had been revised and 13 (4.6%) had undergone exchange of the liner. The most frequent indications for revision were osteolysis and aseptic loosening, followed by polyethylene wear. The mean Kaplan-Meier survival of the Plasmacup was 91% at ten years and 58% at 14 years. Osteolysis was found around 36 (17.1%) of the 211 surviving shells. The median annual rate of linear wear in the surviving shells was 0.12 mm/year and 0.25 mm/year in those which had been revised (p < 0.001). Polyethylene wear was a strong independent risk factor for osteolysis and aseptic loosening. The percentage of patients with osteolysis increased proportionately with each quintile of wear-rate. There is a high late rate of failure of the Plasmacup. Patients with the combination of this prosthesis and bearing should be closely monitored after ten years.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Cementless total hip arthroplasty using a threaded cup and a rectangular tapered stem. Follow-up for ten to 17 years.

We carried out a clinical and radiological review of 103 cementless primary hip arthroplasties with a tapered rectangular grit-blasted titanium press-fit femoral component and a threaded conical titanium acetabular component at a mean follow-up of 14.4 years (10.2 to 17.1). The mean Harris hip score at the last follow-up was 89.2 (32 to 100). No early loosening and no fracture of the implant we...

متن کامل

Unsatisfactory results with the cementless Omnifit acetabular component due to polyethylene and severe osteolysis.

A high incidence of acetabular osteolysis (43%), associated with osteolytic lesions in the proximal femur (22.6%) and leading to a high revision rate, was experienced with the Omnifit total hip prosthesis. We reviewed the clinical and radiological results with 429 Omnifit total hips in 356 patients after a mean follow-up of 60 months. Time to revision and wear of the polyethylene liner with dif...

متن کامل

Anatomic Anteversion of the Acetabular Component Correlates with Polyethylene Linear Wear in Total Hip Arthroplasty: The Three-Dimensional Numerical Analysis

Background: Although abduction of the acetabular component is considered to predict factors for polyethylene wear attributable to osteolysis, other radiographic factors have yet to be elucidated. The purpose of the present study was to evaluate whether anteversion or change in implantation angle of the acetabular component influences polyethylene linear wear by using standing and supine radiogr...

متن کامل

Early Clinical Outcomes with a 3-D Porous Titanium Acetabular Cup

Cementless acetabular components are associated with a significant incidence of polyethylene wear and secondary osteolysis. 3-dimentional (3-D) porous coating and enhanced shell to a liner fixation are expected to reduce the polyethylene wear and to increase the longevity of an acetabular cup. The authors report the early clinical outcome with a cementless acetabular cup, Regenerex Ringloc+ (Zi...

متن کامل

No effect of additional screw fixation of a cementless, all-polyethylene press-fit socket on migration, wear, and clinical outcome

Background and purpose - Additional screw fixation of the all-polyethylene press-fit RM cup (Mathys) has no additional value for migration, in the first 2 years after surgery. However, the medium-term and long-term effects of screw fixation remain unclear. We therefore evaluated the influence of screw fixation on migration, wear, and clinical outcome at 6.5 years using radiostereometric analysi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The Journal of bone and joint surgery. British volume

دوره 92 6  شماره 

صفحات  -

تاریخ انتشار 2010