Dose–volume histogram comparison between static 5-field IMRT with 18-MV X-rays and helical tomotherapy with 6-MV X-rays
نویسندگان
چکیده
We treated prostate cancer patients with static 5-field intensity-modulated radiation therapy (IMRT) using linac 18-MV X-rays or tomotherapy with 6-MV X-rays. As X-ray energies differ, we hypothesized that 18-MV photon IMRT may be better for large patients and tomotherapy may be more suitable for small patients. Thus, we compared dose-volume parameters for the planning target volume (PTV) and organs at risk (OARs) in 59 patients with T1-3 N0M0 prostate cancer who had been treated using 5-field IMRT. For these same patients, tomotherapy plans were also prepared for comparison. In addition, plans of 18 patients who were actually treated with tomotherapy were analyzed. The evaluated parameters were homogeneity indicies and a conformity index for the PTVs, and D2 (dose received by 2% of the PTV in Gy), D98, Dmean and V10-70 Gy (%) for OARs. To evaluate differences by body size, patients with a known body mass index were grouped by that index ( <21; 21-25; and >25 kg/m(2)). For the PTV, all parameters were higher in the tomotherapy plans compared with the 5-field IMRT plans. For the rectum, V10 Gy and V60 Gy were higher, whereas V20 Gy and V30 Gy were lower in the tomotherapy plans. For the bladder, all parameters were higher in the tomotherapy plans. However, both plans were considered clinically acceptable. Similar trends were observed in 18 patients treated with tomotherapy. Obvious trends were not observed for body size. Tomotherapy provides equivalent dose distributions for PTVs and OARs compared with 18-MV 5-field IMRT. Tomotherapy could be used as a substitute for high-energy photon IMRT for prostate cancer regardless of body size.
منابع مشابه
Evaluation of Tumor Control and Normal Tissue Complication Probability in Head and Neck Cancers with Different Sources of Radiation: A Comparative Study
Introduction: The ultimate goal of radiation treatment planning is to yield a high tumor control probability (TCP) with a low normal tissue complication probability (NTCP). Historically dose volume histogram (DVH) with only volumetric dose distribution was utilized as a popular tool for plan evaluation hence present study aimed to compare the radiobiological effectiveness of the cobalt-60 (Co...
متن کاملComparative analysis of 60Co intensity-modulated radiation therapy.
In this study, we perform a scientific comparative analysis of using (60)Co beams in intensity-modulated radiation therapy (IMRT). In particular, we evaluate the treatment plan quality obtained with (i) 6 MV, 18 MV and (60)Co IMRT; (ii) different numbers of static multileaf collimator (MLC) delivered (60)Co beams and (iii) a helical tomotherapy (60)Co beam geometry. We employ a convex fluence m...
متن کاملIntegral radiation dose to normal structures with conformal external beam radiation.
BACKGROUND This study was designed to evaluate the integral dose (ID) received by normal tissue from intensity-modulated radiotherapy (IMRT) for prostate cancer. METHODS AND MATERIALS Twenty-five radiation treatment plans including IMRT using a conventional linac with both 6 MV (6MV-IMRT) and 20 MV (20MV-IMRT), as well as three-dimensional conformal radiotherapy (3DCRT) using 6 MV (6MV-3DCRT)...
متن کاملAbstractID: 3490 Title: Non-Tumor Integral Dose in Conformal, External Beam Radiation Therapy
ID: 3490 Title: Non-Tumor Integral Dose in Conformal, External Beam Radiation Therapy Purpose: To investigate the computed non-tumor integral dose (NTID) delivered during treatment of prostate cancer with intensity modulated radiotherapy (IMRT) compared to three-dimensional conformal radiotherapy (3DCRT). Low and highenergy photon treatments were also evaluated in this study, as was helical tom...
متن کاملThe effect of high energy photon beam in dose- volume parameters and integral dose of normal structures in Head and neck IMRT
Introduction: The aim of this study is to investigate the effect of partially used high photon energies on dosimetric parameters and integral doses of nasopharynx IMRT plans. Materials and Methods: Two 7 coplanar IMRT plan were created for eleven patients (6 MV only IMRT plan in both phase1 and phase2 and 6 MV phase1-mixed energies (6 MV and 15 MV) phase2...