Amperometric Sensor for Detection of Chloride Ions†
نویسندگان
چکیده
Chloride ion sensing is important in many fields such as clinical diagnosis, environmental monitoring and industrial applications. We have measured chloride ions at a carbon paste electrode (CPE) and at a CPE modified with solid AgNO₃, a solution of AgNO₃ and/or solid silver particles. Detection limits (3 S/N) for chloride ions were 100 μM, 100 μM and 10 μM for solid AgNO₃, solution of AgNO₃ and/or solid silver particles, respectively. The CPE modified with silver particles is the most sensitive to the presence chloride ions. After that we approached to the miniaturization of the whole electrochemical instrument. Measurements were carried out on miniaturized instrument consisting of a potentiostat with dimensions 35 × 166 × 125 mm, screen printed electrodes, a peristaltic pump and a PC with control software. Under the most suitable experimental conditions (Britton-Robinson buffer, pH 1.8 and working electrode potential 550 mV) we estimated the limit of detection (3 S/N) as 500 nM.
منابع مشابه
Highly Sensitive Amperometric Sensor Based on Gold Nanoparticles Polyaniline Electrochemically Reduced Graphene Oxide Nanocomposite for Detection of Nitric Oxide
A sensitive electrochemical sensor was fabricated for selective detection of nitric oxide (NO) based on electrochemically reduced graphene (ErGO)-polyaniline (PANI)-gold nanoparticles (AuNPs) nanocomposite. It was coated on a gold (Au) electrode through stepwise electrodeposition to form AuNPs-PANI-ErGO/Au electrode. The AuNPs-PANI-rGO nanocomposite was characterized by Field Emission Scanning ...
متن کاملAMPEROMETRIC MICROBIOSENSOR FOR THE DETERMINATION OF CHOLINE
The construction and performance of an amperometric microbiosensor for the determination of choline chloride by immobilizing choline oxidase on a cellulose acetate membrane are described. The procedure of choline determination involved the electrochemical detection of enzymatically generated hydrogen peroxide. This microsensor has a linear response between 0.038-0.1 mM choline chloride. The...
متن کاملNonenzymatic glucose detection at ordered mesoporous carbon modified electrode.
This work reports on a novel nonenzymatic amperometric glucose sensor based on the ordered mesoporous carbon (OMC). Amperometric method was used to evaluate the electrocatalytic activity of the OMC modified electrode toward nonenzymatic glucose in alkaline media in presence and absence of chloride ions. The results indicated that OMC showed electrocatalytic activity for the oxidation of glucose...
متن کاملPolymeric Membrane Sensor for Potentiometric Determination of Tin(II) Ions
Tin(II)-ion-selective electrode consisting of dibenzo-18-crown-6 (DB18C6) as an ionophore inPVC matrix was developed. The influences of membrane composition, selectivity, response time,reversibility and temperature on the electrode performance were investigated. The electrodeshowed a near Nernstian response for Sn2+ ions over a concentration range from 1.0× 10-6 – 1.0× 10-2 M with a slope of 27...
متن کاملNew Coated Graphite Potentiometric Sensor for Selective Determination of Copper (II) Ions
A highly selective copper (II) coated graphite sensor was prepared by 1,13-bis(8-quinolyl)- 1,4,7,10,13-pentaoxatridecane (kryptofix 5) as a supramolecule ionophore into plasticized polyvinyl chloride (PVC) membrane. The best response characteristic was observed using the membrane composition of PVC = 30.0 mg, dioctyl sebacate (DOS) = 63.5 mg, palmitic acid (PA) = 3.0 mg and kryptofix 5 = 3.5 m...
متن کامل