Evaluation of six commercial kits for detection of human immunoglobulin M antibodies to Toxoplasma gondii. The FDA Toxoplasmosis Ad Hoc Working Group.

نویسندگان

  • M Wilson
  • J S Remington
  • C Clavet
  • G Varney
  • C Press
  • D Ware
چکیده

As a result of reports received by the Food and Drug Administration (FDA) of false-positive results obtained with FDA-cleared in vitro diagnostic kits for the detection of Toxoplasma-specific human immunoglobulin M (IgM) antibodies, an FDA-sponsored evaluation of six kits was performed. A battery of 258 serum specimens, including 30 specimens drawn 1 to 5 months after initial Toxoplasma infection and 228 specimens from Toxoplasma IgG-positive individuals, Toxoplasma IgG-negative individuals, rheumatoid factor-positive persons, and persons determined to be Toxoplasma IgM positive by commercially available assays, was assembled, randomly assorted, and coded. The battery was tested at the FDA with six commercially available kits, at the Palo Alto Medical Foundation (PAMF) by the PAMF double-sandwich IgM enzyme-linked immunosorbent assay (PAMF IgM ELISA), and at the Centers for Disease Control and Prevention (CDC) by the CDC EIA IgM. The results of the PAMF IgM ELISA that were obtained with the battery were considered to be the "gold standard" for this study; specificity rates were computed by considering the PAMF results to be 100% specific. Sensitivity and specificity rates were found to be as follows: CDC EIA IgM, 100 and 99.1%, respectively; Abbott IMx Toxo IgM, version 1, 100 and 77.5%, respectively; Abbott IMx Toxo IgM, version 2, 93.3 and 97.3%, respectively; Abbott Toxo-M EIA, 100 and 84.2%, respectively; BioMérieux Vitek VIDAS Toxo IgM, 100 and 98.6%, respectively; BioWhittaker Toxocap-M, 100 and 95.9%, respectively; Gull Toxo IgM, 97 and 85.6%, respectively; and Sanofi Diagnostics Pasteur Platelia Toxo IgM, 100 and 96.8%, respectively. Although the extent of false-positive reactions with these kits cannot be calculated because the study was retrospective and sample choices were biased, the results may be useful as an indicator of the relative specificities of these kits.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Serological Evaluation of Experimental Toxoplasma gondii Infection in Cats by Using Immunoblotting Based on an Affinity Purified Surface Antigen

Toxoplasma gondii is an apicomplexan parasite that infects human and almost all warm-blooded animals. The life cycle of the parasite includes an asexual reproduction in intermediate hosts (Mammals and birds) and a sexual reproduction in definitive hosts (Felidae). Cats are both the intermediate and the definitive host for T. gondii. The aim of this study was to investigate anti-T. gondii antibo...

متن کامل

Detection of immunoglobulin G (IgG) and IgM antibodies to Toxoplasma gondii: evaluation of four commercial immunoassay systems.

A comparative evaluation of the following commercial immunoassays for the determination of antibodies to Toxoplasma gondii was performed: Behring Diagnostics OPUS Toxo G and Toxo M, Abbott Diagnostics IMX Toxo-IgG 2.0 and Toxo-IgM, Sanofi Diagnostics Pasteur Platelia Toxo IgG and Toxo IgM, and bioMérieux Vitek VIDAS Toxo IgG and IgM. Of 676 specimens that were tested for Toxoplasma-specific imm...

متن کامل

Electrochemiluminesence Epidemiologic Detection of Toxoplasma gondii Infection in Pregnant Women With Direct and Indirect Diagnostic Techniques (ELISA Avidity Plus Biochemical Assay)

Background and aims: Toxoplasmosis is a very common disease in the world. Two types of acquiredtoxoplasmosis have been detected. In the chronic toxoplasmosis, the abnormality of tissue functionis little but, in acute toxoplasmosis, function of RES system becomes interrupted. The assessment oftoxoplasma antibody with tissue enzymes in this stage is very important. Furthermore, ...

متن کامل

Interlaboratory evaluation of indirect enzyme-linked immunosorbent assay, antibody capture enzyme-linked immunosorbent assay, and immunoblotting for detection of immunoglobulin M antibodies to Toxoplasma gondii.

One hundred and fifteen serum samples from healthy laboratory personnel and 50 consecutive samples from 19 patients with anamnestic clinical signs of toxoplasmosis were assayed by four laboratories for the presence of immunoglobulin M antibodies to Toxoplasma gondii by an indirect enzyme-linked immunosorbent assay (ELISA), an antibody capture assay with peroxidase-labeled toxoplasma antigen, an...

متن کامل

ارزیابی بین آزمایشگاهی و بین کیتی آزمون الایزا برای تشخیص آنتی بادی های IgG اختصاصی علیه توکسوپلاسما گوندی ای

Background and purpose: Toxoplasmosis is a parasitic protozoan disease producing severe complications in congenital and immunocompromised cases. The enzyme-linked immunosorbent assay (ËLÏSÂ) is the most common laboratory method for serodiagnosis of toxoplasmosis. Çommercial kits have no unique procedure for standardization and this may influence the consistency of the results. The present stu...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of clinical microbiology

دوره 35 12  شماره 

صفحات  -

تاریخ انتشار 1997