Radiation therapy treatments have become increasingly more complicated. There are multiple opportunities for humans, machines, software, and combinations thereof to result in a treatment error that could be of significance. Current methods for quality assurance are often abstract in nature and may have unclear underlying assumptions as to what is assumed to be working correctly, or may depend upon the diligence of persons to discover errors from a review of the treatment plan. Here, an example will be shown of a direct method to reconstruct and demonstrate the dose and the dose distribution delivered to a particular patient. By measuring the radiation fields that come out of the accelerator, and using the measurement as input to a 3-dimensional (3D) dose algorithm, the delivered patient dose is determined and presented in a manner similar to the treatment plan. The intended treatment plan dose may be directly compared. Using this feedback mechanism, there is less abstraction and dependence upon the diligence of individuals checking multiple steps in a treatment process, and assumptions can be clearly stated. With this system, the dose is determined and presented minimizing assumptions and dependence upon other systems.
منابع مشابه
3D dose reconstruction in external beam radiotherapy using portal imaging
The verification of the dose delivered to a patient during radiation therapy is a particularly important step for ensuring treatment quality. This step, known as in vivo dosimetry can be realized using a bidimensional dosimeter named Electronic Portal Imaging Device (EPID). However, few methods enable the determination of an accurate and fully personalized 3D dose distribution for the most curr...
متن کاملEvaluation of Accuracy and Quality assurance of external beam therapy with photons
Introduction: Receiving exact dose by the patients is vital in radiotherapy. In radiation therapy, the dosimetry of radiations is too important because of successful radiation inquires for delivering the exact dose to the target volume. This study is to evaluate the tolerances and the accuracy of calculated dose of photon beams in the treatment software system. The TECDOC1583 p...
متن کاملDosimetric Comparison of 3D Tangential Radiotherapy of Post-Lumpectomy Breast at Two Different Energies
Introduction: Radiation therapy following breast conserving surgery is one of the most common procedures performed in any radiation oncology department. A tangential parallel-opposed pair is almost always the technique of choice for this purpose. This technique is often performed based on 3D treatment planning. The aim of this study was to compare 3D treatment planning for two different energie...
متن کامل3D-Conformal Radiation Therapy and Intensity-Modulated Radiation Therapy Techniques for Laryngeal Cancer Taking Parotid Glands as Organ at Risk
Background: Three-dimensional conformal radiotherapy (3D-CRT) and intensity-modulated radiotherapy (IMRT) techniques are used for the treatment of patients with laryngeal cancer. Objective: This study aimed to investigate the effects of these 2 treatment techniques on the planning target volume (PTV) (laryngeal cancer), dose homogeneity, dose of organs at...
متن کاملExternal radiotherapy of intact breast: A comparison between 2D (single CT- slice) and 3D (full CT-slices) plans
Background: Tangential irradiation of intact breast is one of the most common procedures performed in any radiotherapy center. This method is performed by using 2D and 3D treatment planning. The aim of this study was to compare 2D with 3D plans in breast conserving radiotherapy. Homogeneity of isodose, and lung received dose were compared. Materials and Methods: Twenty patients with br...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Medical dosimetry : official journal of the American Association of Medical Dosimetrists
دوره 32 3 شماره
صفحات -
تاریخ انتشار 2007