Experimental validation of Doppler echocardiographic measurement of volume flow through the stenotic aortic valve.
نویسندگان
چکیده
In aortic stenosis, evaluation of aortic valve area by the continuity equation assumes that the volume of flow through the stenotic valve can be measured accurately in the left ventricular outflow tract. To test the accuracy of Doppler volume-flow measurement proximal to a stenotic valve, we developed an open-chest canine model in which the native leaflets were sutured together to create variable degrees of acute aortic stenosis. Left ventricular and aortic pressures were measured with micromanometer-tipped catheters. Volume flow was controlled and varied by directing systemic venous return through a calibrated roller pump and back to the right atrium. Because transaortic volume flow will not equal roller pump output when there is coexisting aortic insufficiency (present in 67% of studies), transaortic flow was measured by electromagnetic flowmeter with the flow probe placed around the proximal descending thoracic aorta, just beyond the ligated arch vessels. In 12 adult, mongrel dogs (mean weight, 25 kg), the mean transaortic pressure gradient ranged from 2 to 74 mm Hg, and transaortic volume flow ranged from 0.9 to 3.2 l/min. In four dogs, electromagnetic flow that was measured distal to the valve was accurate compared with volume flow determined by timed collection of total aortic flow into a graduated cylinder (n = 24, r = 0.97, electromagnetic flow = 0.87 Direct +0.13 l/min). In eight subsequent dogs, electromagnetic flow was compared with transaortic cardiac output measured by Doppler echocardiography in the left ventricular outflow tract as circular cross-sectional area [pi(D/2)2] x left ventricular outflow tract velocity-time integral x heart rate.(ABSTRACT TRUNCATED AT 250 WORDS)
منابع مشابه
Stenotic Aortic Valve Area
I n aortic stenosis, the aortic valve area (AVA) is intuitively the most appealing measure of severity because it is supposed to be (largely) independent of stroke volume and left ventricular pump function and reflects most closely “stenosis.” Calculation of the AVA by the continuity equation is routinely performed in echocardiography (1), using 3 measurements: the velocity-time integral of the...
متن کاملVariation of anatomic valve area during ejection in patients with valvular aortic stenosis evaluated by two-dimensional echocardiographic planimetry: comparison with traditional Doppler data.
OBJECTIVES Flow variations can affect valve-area calculation in aortic stenosis and lead to inaccuracies in the evaluation of the stenosis. Knowing that transvalvular flow varies normally within one beat, we designed this study to assess the response of the valve to intrabeat variation of flow during systole. Results were compared with flow-derived measurements. BACKGROUND Technological impro...
متن کاملDoppler echocardiographic findings in tissue engineered aortic valve in a sheep model
Background: Heart valve diseases are considered a common disease in human and animals, and valve replacement is an option for treatment of valvular diseases. Objectives: In this study efficacy of a tissue engineered valve in thoracic aorta was evaluated with transthoracic echocardiography. Methods: This study was undertaken on 6 male sheep. Echocardiography was performed on all sheep 24 hours b...
متن کاملA new non-invasive estimation of the stenotic aortic valve area by pulsed Doppler mapping.
A new pulsed Doppler mapping technique has been used to measure the severity of aortic valve stenosis. The Doppler examination was performed at the site of the aortic orifice in the parasternal short axis echocardiographic view and the method was based on the detection of the area of systolic flow through the stenotic orifice. This area was derived by planimetry and the measurements obtained by...
متن کاملA new approach for the evaluation of the severity of coarctation of the aorta using Doppler velocity index and effective orifice area: in vitro validation and clinical implications.
Early detection and accurate estimation of COA severity are the most important predictors of successful long-term outcome. However, current clinical parameters used for the evaluation of the severity of COA have several limitations and are flow dependent. The objectives of this study are to evaluate the limitations of current existing parameters for the evaluation of the severity of coarctation...