Patient-specific quality assurance of RapidArc treatments: Portal prediction dosimetry compared with phantom studies
نویسنده
چکیده
Purpose: To validate a locally fabricated phantom of Imatrixx-2D Array by comparing its results with ArcCheck phantom and comparing portal dosimetry measurements with the two phantom studies. Materials and Methods: Electronic Portal Imaging Devices and Epiqa software were used for portal dosimetry. An Imatrixx-2D array with a locally fabricated phantom and ArcCheck cylindrical phantom were used for phantom studies. Eclipse-TPS with RapidArc treatment planning and portal dose prediction software was used for planar dose calculations. Three verification plans were created for each of the 15 patient plans of various sites, making a total of 45 plans to be delivered on 3 QA systems as above. Fifteen plans each with 2 arcs were delivered on the EPIDs of the Linacs, on Imatrixx-2D array phantom and on ArcCheck cylindrical phantom respectively. The planar dose matrices were analysed using global Gamma Index criteria of 3mm DTA and 3% dose difference. Results: The maximum deviations of percentage in dose points, in which γ>1, are 1.94, 1.89 and 1.5 in Imatrixx phantom, ArcCheck phantom and Portal dosimetry, respectively. Similarly, the mean deviations and SD values are less in portal dosimetry than that of phantom studies. The smaller deviations in portal dosimetry are attributed to closely embedded chambers in the EPID compared to the distance between the detectors placed in the phantom measurements. Conclusion: After carrying out the comparison of results, the locally fabricated phantom has been validated and accepted for the dosimetric studies. The conclusion is that all the three dosimetric QA systems are suitable for the patient-specific QA of RapidArc treatments. © 2012 Biomedical Imaging and Intervention Journal. All rights reserved.
منابع مشابه
A patient-specific quality assurance study on absolute dose verification using ionization chambers of different volumes in RapidArc treatments.
The recalculation of 1 fraction from a patient treatment plan on a phantom and subsequent measurements have become the norms for measurement-based verification, which combines the quality assurance recommendations that deal with the treatment planning system and the beam delivery system. This type of evaluation has prompted attention to measurement equipment and techniques. Ionization chambers ...
متن کاملA quality assurance program for an amorphous silicon electronic portal imaging device using in-house developed phantoms: a method development for dosimetry purposes
Background: Electronic portal imaging devices (EPIDs) play an important role in radiation therapy portal imaging, geometric and dosimetric verifications. A successful utilization of EPIDs for imaging and dosimetric purposes requires a reliable quality control process routine to be carried out regularly. In this study, two in-house phantoms were developed and analyzed for implementation in a qua...
متن کاملCharacterization of responses of 2d array seven29 detector and its combined use with octavius phantom for the patient-specific quality assurance in rapidarc treatment delivery.
A commercial 2D array seven29 detector has been characterized and its performance has been evaluated. 2D array ionization chamber equipped with 729 ionization chambers uniformly arranged in a 27 × 27 matrix with an active area of 27 × 27 cm² was used for the study. An octagon-shaped phantom (Octavius Phantom) with a central cavity is used to insert the 2D ion chamber array. All measurements wer...
متن کاملA Specific Patient Quality Assurance (PSQA) procedure for a Co-60 source based High Dose Rate Brachytherapy
Introduction: In radiation therapy, accurate dose determination and precise dose delivery to the tumor are directly associated with better treatment outcomes in terms of higher tumor control and lower post radiation therapy complications. The current study aims the development and clinical application of the Patient Specific Quality Assurance (PSQA) procedures for nasopharyngea...
متن کاملIn Vivo Dosimetry Using a Flat Surface Sun Nuclear Corporation Diode in 60co Beams for Some Radiotherapy Treatments in Ghana
Introduction: One of the useful standard quality assurance techniques in radiation therapy is monitoring entrance doses in in-vivo dosimetry. An overall tolerance limit of 5% of the absorbed radiation dose has been recommended by the International Commission of Radiological Units. The implementation of an in vivo dosimet...