Dose optimization of breast balloon brachytherapy using a stepping Ir192 HDR source

نویسندگان

  • Chang Hyun Choi
  • Sung‐Joon Ye
  • E. Ishmael Parsai
  • Sui Shen
  • Ruby Meredith
  • Ivan A. Brezovich
  • Roger Ove
چکیده

There is a considerable underdosage (11%-13%) of PTV due to anisotropy of a stationary source in breast balloon brachytherapy. We improved the PTV coverage by varying multiple dwell positions and weights. We assumed that the diameter of spherical balloons varied from 4.0 cm to 5.0 cm, that the PTV was a 1-cm thick spherical shell over the balloon (reduced by the small portion occupied by the catheter path), and that the number of dwell positions varied from 2 to 13 with 0.25-cm steps, oriented symmetrically with respect to the balloon center. By assuming that the perfect PTV coverage can be achieved by spherical dose distributions from an isotropic source, we developed an optimization program to minimize two objective functions defined as: (1) the number of PTV-voxels having more than 10% difference between optimized doses and spherical doses, and (2) the difference between optimized doses and spherical doses per PTV-voxel. The optimal PTV coverage occurred when applying 8-11 dwell positions with weights determined by the optimization scheme. Since the optimization yields ellipsoidal isodose distributions along the catheter, there is relative skin sparing for cases with source movement approximately tangent to the skin. We also verified the optimization in CT-based treatment planning systems. Our volumetric dose optimization for PTV coverage showed close agreement to linear or multiple-points optimization results from the literature. The optimization scheme provides a simple and practical solution applicable to the clinic.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Consideration of accurate dose distribution for Contura balloon catheter in breast brachytherapy with MCNP5 simulation

Introduction: Contura is a balloon applicator that used to deliver intra cavity radiation for Partial Breast Irradiation (PBI) treatments. Contura with 4 peripheral lumens surrounding the central, for loading HDR sources from an after loader system can provide an asymmetric dose distribution and gives a more flexible shape of dose than earlier balloon applicators such as Mammo...

متن کامل

Perioperative high dose rate (HDR) brachytherapy in unresectable locally advanced pancreatic tumors

PURPOSE The aim of the study was to present an original technique of catheter implantation for perioperative HDR-Ir192 brachytherapy in patients after palliative operations of unresectable locally advanced pancreatic tumors and to estimate the influence of perioperative HDR-Ir192 brachytherapy on pain relief in terminal pancreatic cancer patients. MATERIAL AND METHODS Eight patients with panc...

متن کامل

A Comparison of Skin Dose Delivered with MammoSite and Multicatheter Breast Brachytherapy

Background: Accelerated partial breast irradiation via interstitial balloon brachytherapy is a fast and effective treatment method for certain early stage breast cancers however skin, chest wall and Lung doses are correlated with toxicity in patients treated with breast brachytherapy.Objective: To investigate the percentage of the dose received by critical organ (skin), thermoluminescence detec...

متن کامل

Calculation of absorbed dose rate in the lungs, ribs and skin by the mammosite applicator in breast cancer brachytherapy with MCNP

Introduction: Breast cancer is the most common cancer among women. Brachytherapy is one of the ways to treat breast cancer. In brachytherapy treatments, high-dose sources are used in interstitial placement. Here, Iridium-192 is used. This source emits gamma rays with energies ranging from 136 kV to 1060 kV electron volts. The activity of the used spring is 10 Ci. The simulated ...

متن کامل

A study on dosimetry accuracy of Strut-Adjust Volume Implant (SAVI) brachytherapy

Introduction: Accelerated Partial Breast Irradiation (APBI) is an effective treatment for breast carcinomas because of its limited number of fractions. Strut-adjusted volume implant (SAVI) is a HDR brachytherapy applicator to deliver the prescription dose with good PTV coverage and acceptable dose to skin, chest wall and organ at risks. Acceptable clinical outcomes depend on a...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 10  شماره 

صفحات  -

تاریخ انتشار 2009