Field effect transistor with integrated microfluidic channel as pH sensor
نویسندگان
چکیده
This paper presents an original design of chemical sensors with an integrated microfluidic channel. Targeted applications are pH-meters devices. The integration of the microfluidic channel allows decreasing the volume required for each measurement. The sensing part of the device consists of a field effect transistor with a suspended gate directly performed above the fluidic channel. Chemicals under test are driven through the sensing area between the electrical channel of the FET and the suspended gate. By this way products that flow in the microfluidic channel directly module the concentration of charges inside the transistor’s gap and thus induce changes in the transfer characteristic. This paper describes the fabrication process and the technological choices for materials. Electrical tests, performed in air and in liquid, have shown a good behavior of the transistor, linked to a good mechanical sustain of the fluidic channel. The system is able to detect transition between air and liquid media. Moreover, it has shown a high sensitivity (about 300 mV/pH) to pH measurements.
منابع مشابه
Detection of Micrococcus Luteus Biofilm Formation in Microfluidic Environments by pH Measurement Using an Ion-Sensitive Field-Effect Transistor
Biofilm formation in microfluidic channels is difficult to detect because sampling volumes are too small for conventional turbidity measurements. To detect biofilm formation, we used an ion-sensitive field-effect transistor (ISFET) measurement system to measure pH changes in small volumes of bacterial suspension. Cells of Micrococcus luteus (M. luteus) were cultured in polystyrene (PS) microtub...
متن کاملIntegrated single-walled carbon nanotube/microfluidic devices for the study of the sensing mechanism of nanotube sensors.
A method to fabricate integrated single-walled carbon nanotube/microfluidic devices was developed. This simple process could be used to directly prepare nanotube thin film transistors within the microfluidic channel and to register SWNT devices with the microfludic channel without the need of an additional alignment step. The microfluidic device was designed to have several inlets that deliver ...
متن کاملIntegration of a Droplet-Based Microfluidic System and Silicon Nanoribbon FET Sensor
We present a novel microfluidic system that integrates droplet microfluidics with a silicon nanoribbon field-effect transistor (SiNR FET), and utilize this integrated system to sense differences in pH. The device allows for selective droplet transfer to a continuous water phase, actuated by dielectrophoresis, and subsequent detection of the pH level in the retrieved droplets by SiNR FETs on an ...
متن کاملEffect of Aspect Ratio (Length:Diameter) on a Single Polypyrrole Nanowire FET Device
The effect of different aspect ratios (length to diameter ratio, L:D) on single polypyrrole (Ppy) nanowire based field effect transistor (FET) sensors for real time pH monitoring was studied. Ppy nanowires with diameters of ∼60, ∼80, and ∼200 nm were synthesized by using electrochemical deposition inside anodized aluminum oxide (AAO) template and were assembled using AC dielectrophoretic alignm...
متن کاملAn integrated silicon sensor with microfluidic chip for monitoring potassium and pH
We present ion-sensitive field effect transistor-based sensors, integrated with a microfluidic chip, for monitoring pH and potassium cations. The sensor is stra-tegically located at the base of a well so that the response time of the device depends both on the mean low through the device and the diffusion coefficient of the analyte being monitored. This would enable monitoring of ions in the pr...
متن کامل