Radiotherapy dosimetry using a commercial OSL system.
نویسندگان
چکیده
A commercial optically stimulated luminescence (OSL) system developed for radiation protection dosimetry by Landauer, Inc., the InLight microStar reader, was tested for dosimetry procedures in radiotherapy. The system uses carbon-doped aluminum oxide, Al2O3:C, as a radiation detector material. Using this OSL system, a percent depth dose curve for 60Co gamma radiation was measured in solid water. Field size and SSD dependences of the detector response were also evaluated. The dose response relationship was investigated between 25 and 400 cGy. The decay of the response with time following irradiation and the energy dependence of the Al2O3:C OSL detectors were also measured. The results obtained using OSL dosimeters show good agreement with ionization chamber and diode measurements carried out under the same conditions. Reproducibility studies show that the response of the OSL system to repeated exposures is 2.5% (1sd), indicating a real possibility of applying the Landauer OSL commercial system for radiotherapy dosimetric procedures.
منابع مشابه
Optically stimulated luminescence in vivo dosimetry for radiotherapy: physical characterization and clinical measurements in (60)Co beams.
A commercial optically stimulated luminescence (OSL) dosimetry system was investigated for in vivo dosimetry in radiation therapy. Dosimetric characteristics of InLight dot dosimeters and a microStar reader (Landauer Inc.) were tested in (60)Co beams. The reading uncertainty of a single dosimeter was 0.6%. The reproducibility of a set of dosimeters after a single irradiation was 1.6%, while in ...
متن کاملOptically stimulated luminescence (OSL) of carbon-doped aluminum oxide (Al2O3:C) for film dosimetry in radiotherapy.
INTRODUCTION AND PURPOSE Conventional x-ray films and radiochromic films have inherent challenges for high precision radiotherapy dosimetry. Here we have investigated basic characteristics of optically stimulated luminescence (OSL) of irradiated films containing carbon-doped aluminum oxide (Al2O3:C) for dosimetry in therapeutic photon and electron beams. MATERIALS AND METHODS The OSL films co...
متن کاملAdvantages and Disadvantages in the Utilisation of Thermoluminescence (tl) and Optically Stimulated Luminescence (osl) for Radiation Dosimetry
Thermoluminescence is an established method for radiation dosimetry as well as retrospective dosimetry. In spite of its great success, different difficulties are associated with its application. These include anomalous fading, changes in the properties of the sample due to its heating in the readout phase, including a sensitization due to the combined effect of irradiation and heating and non-l...
متن کاملDevelopment of Osl Reader Systems for Radiation Dosimetry
Thermoluminescence (TL) and optically stimulated luminescence (OSL) techniques are known as a research tools for the study of phosphor materials and for their applications in radiation dosimetry in particular [1]. Although TL technique is known since several decades, OSL is relatively newer technique. The TL or OSL emission from an irradiated sample arises upon stimulation with heat or light re...
متن کاملStudy of agate using the OSL technique and application in
The possibility of using Brazilian stones in gamma high-dose dosimetry was investigated at IPEN using the optically stimulated luminescence (OSL) technique. Agate samples were studied in relation to their OSL dosimetric properties, in this work. The samples were irradiated at a Gamma-Cell system (Co). The calibration curves show a sublinear behavior between 50 Gy and 300 kGy, and the lower dete...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Medical physics
دوره 35 4 شماره
صفحات -
تاریخ انتشار 2008