No effect of additional screw fixation of a cementless, all-polyethylene press-fit socket on migration, wear, and clinical outcome

نویسندگان

  • Michiel J M Minten
  • Petra J C Heesterbeek
  • Maarten Spruit
چکیده

Background and purpose - Additional screw fixation of the all-polyethylene press-fit RM cup (Mathys) has no additional value for migration, in the first 2 years after surgery. However, the medium-term and long-term effects of screw fixation remain unclear. We therefore evaluated the influence of screw fixation on migration, wear, and clinical outcome at 6.5 years using radiostereometric analysis (RSA). Patients and methods - This study involved prolonged follow-up from a previous randomized controlled trial (RCT). We analyzed RSA radiographs taken at baseline and at 1-, 2-, and 6.5-year follow-up. Cup migration and wear were assessed using model-based RSA software. Wear was calculated as translation of the femoral head model in relation to the cup model. Total translation, rotation, and wear were calculated mathematically from results of the orthogonal components. Results - 27 patients (15 with screw fixation and 12 without) were available for follow-up at 6.5 (5.6-7.2) years. Total translation (0.50 mm vs. 0.56 mm) and rotation (1.01 degrees vs. 1.33 degrees) of the cup was low, and was not significantly different between the 2 groups. Wear increased over time, and was similar between the 2 groups (0.58 mm vs. 0.53 mm). Wear rate (0.08 mm/year vs. 0.09 mm/year) and clinical outcomes were also similar. Interpretation - Our results indicate that additional screw fixation of all-polyethylene press-fit RM cups has no additional value regarding medium-term migration and clinical outcome. The wear rate was low in both groups.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A cementless, elastic press-fit socket with and without screws

BACKGROUND The acetabular component has remained the weakest link in hip arthroplasty regarding achievement of long-term survival. Primary fixation is a prerequisite for long-term performance. For this reason, we investigated the stability of a unique cementless titanium-coated elastic monoblock socket and the influence of supplementary screw fixation. PATIENT AND METHODS During 2006-2008, we...

متن کامل

Cementless hemispheric porous-coated sockets implanted with press-fit technique without screws: average ten-year follow-up.

BACKGROUND Press-fit implantation of a porous-coated hemispheric acetabular component without screws is an option for primary total hip replacement. The purpose of the present study was to evaluate the results of this technique after an average duration of follow-up of ten years to determine if there was any loss of fixation or increase in osteolysis over time. METHODS From June 1988 to Novem...

متن کامل

Superior fixation of pegged trabecular metal over screw-fixed pegged porous titanium fiber mesh

BACKGROUND AND PURPOSE Lasting stability of cementless implants depends on osseointegration into the implant surface, and long-term implant fixation can be predicted using radiostereometric analysis (RSA) with short-term follow-up. We hypothesized that there would be improved fixation of high-porosity trabecular metal (TM) tibial components compared to low-porosity titanium pegged porous fiber-...

متن کامل

Unsatisfactory results with the cementless Omnifit acetabular component due to polyethylene and severe osteolysis.

A high incidence of acetabular osteolysis (43%), associated with osteolytic lesions in the proximal femur (22.6%) and leading to a high revision rate, was experienced with the Omnifit total hip prosthesis. We reviewed the clinical and radiological results with 429 Omnifit total hips in 356 patients after a mean follow-up of 60 months. Time to revision and wear of the polyethylene liner with dif...

متن کامل

Fixation, polyethylene wear, and pelvic osteolysis in primary total hip replacement.

A multicenter retrospective review was performed analyzing 1081 primary total hip replacements in 944 patients using the Harris Galante-I cementless acetabular component with screw fixation. All patients were followed up for a minimum of 5 years with a mean followup of 81 months. Linear polyethylene wear averaged 0.11 mm/year (range, 0-0.86 mm/year). Pelvic osteolysis was seen in 25 patients (2...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 87  شماره 

صفحات  -

تاریخ انتشار 2016