Wear Distribution Detection of Knee Joint Prostheses by Means of 3D Optical Scanners
نویسندگان
چکیده
The objective of this study was to examine total knee polyethylene inserts from in vitro simulation to evaluate and display-using a 3D optical scanner-wear patterns and wear rates of inserts exposed to wear by means of simulators. Various sets of tibial inserts have been reconstructed by using optical scanners. With this in mind, the wear behavior of fixed and mobile bearing polyethylene knee configurations was investigated using a knee wear joint simulator. After the completion of the wear test, the polyethylene menisci were analyzed by an innovative 3D optical scanners in order to evaluate the 3D wear distribution on the prosthesis surface. This study implemented a new procedure for evaluating polyethylene bearings of joint prostheses obtained after in vitro wear tests and the proposed new approach allowed quantification of the contact zone on the geometry of total knee prostheses. The results of the present study showed that mobile TKPs (total knee prosthesis) have lower wear resistance with respect to fixed TKPs.
منابع مشابه
New Challenges in Tribology: Wear Assessment Using 3D Optical Scanners
Wear is a significant mechanical and clinical problem. To acquire further knowledge on the tribological phenomena that involve freeform mechanical components or medical prostheses, wear tests are performed on biomedical and industrial materials in order to solve or reduce failures or malfunctions due to material loss. Scientific and technological advances in the field of optical scanning allow ...
متن کاملComparison of polyethylene wear in anatomical and reversed shoulder prostheses.
Wear of polyethylene is associated with aseptic loosening of orthopaedic implants and has been observed in hip and knee prostheses and anatomical implants for the shoulder. The reversed shoulder prostheses have not been assessed as yet. We investigated the volumetric polyethylene wear of the reversed and anatomical Aequalis shoulder prostheses using a mathematical musculoskeletal model. Movemen...
متن کاملInvestigating the Impact of Knee Prosthesis in Patients’ Body on Radiation Dose Distribution: A Monte Carlo Approach
Introduction: Metal prostheses in patients affect the radiotherapy dose distribution. Metal prostheses with high density and atomic number cause major changes in scattering and attenuation of radiation. The present study aims to assess the impact of metal knee prosthesis with various dimensions and materials on radiotherapy dose distribution. Material and Methods: In this research, the Varian L...
متن کاملROLLING CONTACT ORTHOPAEDIC JOINT DESIGN by
Arthroplasty, the practice of rebuilding diseased biological joints using engineering materials, is often used to treat severe arthritis of the knee and hip. Prosthetic joints have been created in a “biomimetic” manner to reconstruct the shape of the biological joint. We are at a disadvantage, however, in that metals and polymers used to replace bone and articular cartilage often wear out too s...
متن کاملSimulation of polyethylene wear in ankle joint prostheses.
The performance of total ankle replacements (TARs) have not been comparable to those of the other major joints of the lower extremity. The aim of this work was to develop a new simulator test to compare the wear of a new mobile bearing TAR (Mobility) with one with a good clinical history, the Buechel Pappas, using kinematic inputs derived from the literature. The wear rate for the Mobility comp...
متن کامل