Oximetry of retinal vessels by dual-wavelength imaging: calibration and influence of pigmentation.
نویسندگان
چکیده
A method for noninvasive measurement of Hb O2 saturation (SO2) in retinal blood vessels by digital imaging was developed and tested. Images of vessels were recorded at O2-sensitive and O2-insensitive wavelengths (600 and 569 nm, respectively) by using a modified fundus camera with an image splitter coupled to an 18-bit digital camera. Retinal arterial SO2 was varied experimentally by having subjects breathe mixtures of O2 and N2 while systemic arterial SO2 was monitored with a pulse oximeter. Optical densities (ODs) of vascular segments were determined using a computer algorithm to track the path of reflected light intensity along vessels. During graded hypoxia the OD ratio (ODR = OD600/OD569) bore an inverse linear relationship to systemic SO2. Compensation for the influence of choroidal pigmentation significantly reduced variation in the arterial SO2 measurements among subjects. An O2 sensitivity of 0.00504 +/- 0.00029 (SE) ODR units/%SO2 was determined. Retinal venous SO2 at normoxia was 55 +/- 3.38% (SE). Breathing 100% O2 increased venous SO2 by 19.2 +/- 2.9%. This technique, when combined with blood flow studies in human subjects, will enable the study of retinal O2 utilization under experimental and various disease conditions.
منابع مشابه
Retinal Oximetry and Vessel Diameter Measurements With a Commercially Available Scanning Laser Ophthalmoscope in Diabetic Retinopathy
Purpose To test the hypothesis that retinal vascular diameter and hemoglobin oxygen saturation alterations, according to stages of diabetic retinopathy (DR), are discernible with a commercially available scanning laser ophthalmoscope (SLO). Methods One hundred eighty-one subjects with no diabetes (No DM), diabetes with no DR (No DR), nonproliferative DR (NPDR), or proliferative DR (PDR, all h...
متن کاملImproving the two-wavelength fundus oximetry method for quantitatively measuring retinal oxygen saturation in eye model phantom using a steady state diffusion algorithm
This article has no abstract.
متن کاملRetinal oximetry with a scanning laser ophthalmoscope.
PURPOSE The purpose of the study was to assess if a scanning laser ophthalmoscope (SLO), Optomap 200Tx, could be used for measurements of hemoglobin oxygen saturation in retinal blood vessels. METHODS Optomap 200Tx uses two lasers for image acquisition, 532 and 633 nm. Retinal images of healthy individuals and patients with retinal vein occlusion were analyzed with modified Oxymap Analyzer so...
متن کاملRetinal Oximetry with Scanning Laser Ophthalmoscope in Infants
PURPOSE Dual wavelength retinal oximetry has been developed for adults, but is not available for infants. Retinal oximetry may provide insight into the pathophysiology of oxygen-mediated diseases like retinopathy of prematurity. More insight in the oxygen metabolism of the retina in infants may provide valuable clues for better understanding and subsequent prevention or treatment of the disease...
متن کاملThe effect of simulated cataract light scatter on retinal vessel oximetry.
To assess the impact of light scatter, similar to that introduced by cataract on retinal vessel blood oxygen saturation measurements using poly-bead solutions of varying concentrations. Eight healthy, young, non-smoking individuals were enrolled for this study. All subjects underwent digital blood pressure measurements, assessment of non-contact intraocular pressure, pupil dilation and retinal ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of applied physiology
دوره 86 2 شماره
صفحات -
تاریخ انتشار 1999