A multicenter clinical evaluation of the Clot Signature Analyzer.
نویسندگان
چکیده
BACKGROUND The Clot Signature Analyzer (CSA) was designed to assess global hemostasis as a screening assay using non-anticoagulated whole blood. Three different measurements are produced by the instrument: platelet hemostasis time (PHT), clot time (CT), and collagen-induced thrombus formation (CITF). OBJECTIVES The purpose of the present study was to determine normal ranges for these measurements and assess the performance of the CSA in patients with well-characterized hemostatic disorders and in normal subjects. PATIENTS AND METHODS Four institutions participated in the study. Each established their own normal reference ranges. Patients with well-characterized hemostatic disorders and concurrent normal controls were subsequently examined. RESULTS Normal ranges between institutions were similar although statistically different. One hundred and eight patients were examined: 46 individuals with von Willebrand disease (VWD) (type 1, 26; type 2A, 11; type 2B, six; type 3, three); 38 patients with a coagulation factor deficiency; 13 individuals with platelet function defects; 10 patients taking warfarin; and one individual on low-molecular-weight heparin. Of these patients, 89% had at least one abnormality by CSA: 42/46 VWD patients, 35/38 coagulation protein defect patients, 9/13 patients with platelet function defects, 9/10 patients on warfarin and 1/1 patient on low-molecular-weight heparin. Of 116 normal subjects, 103 (89%) fell within the centers' normal range. These data suggest that the CSA has a good sensitivity for bleeding disorders.
منابع مشابه
Evaluation of thrombolytic potential of three medicinal plants available in Bangladesh, as a potent source of thrombolytic compounds
Objective: The present study is aimed to investigate in vitro thrombolytic activity of three Bangladeshi medicinal plants Averrhoa bilimbi (Oxalidiaceae), Clerodendrum viscosum (Verbanaceae) and Drynaria quercifolia (Polypodiaceae). Materials and methods: Each the plant was extracted with methanol at room temperature and the concentrated methanolic extracts (MEF) were fractionated by the modifi...
متن کاملThe new protocol blind digital signature based on the discrete logarithm problem on elliptic curve
In recent years it has been trying that with regard to the question of computational complexity of discrete logarithm more strength and less in the elliptic curve than other hard issues, applications such as elliptic curve cryptography, a blind digital signature method, other methods such as encryption replacement DLP. In this paper, a new blind digital signature scheme based on elliptic curve...
متن کاملEvaluation of the SonoClot Analyzer for the measurement of platelet function in whole blood.
A viscoelastometer, the SonoClot Coagulation Analyzer, has been proposed for use in the evaluation of platelet function. The purpose of this study was to evaluate systematically this instrument when used with whole blood. Under laboratory conditions, the coefficient of variation (cv) of determinations of whole blood activated clotting time (ACT) on the instrument was approximately 7.0 percent. ...
متن کاملA multicenter evaluation of the Iris iQ200 automated urine microscopy analyzer body fluids module and comparison with hemacytometer cell counts.
Manual hemacytometer cell counting of body fluids is labor-intensive and requires skilled testing personnel. We performed a multicenter evaluation of the Iris iQ200 automated microscopy analyzer Body Fluids Module (Iris Diagnostics, Chatsworth, CA) and compared 350 iQ200 body fluid cell counts with manual hemacytometer cell counts. Within-run imprecision, expressed as coefficient of variation (...
متن کاملMulticenter evaluation of Tosoh glycohemoglobin analyzer.
BACKGROUND We describe an Anglo-French evaluation of a new analyzer. METHODS The Tosoh HLC-723 GHb V, A1c2.2 glycohemoglobin analyzer is an HPLC instrument with primary blood tube sampling, bar-code reading, cap piercing, and the ability to chromatographically separate labile hemoglobin A1c (HbA1c). We evaluated two analytical protocols, 2.2 and 3.0 min, and compared results for blood samples...