Fast inversion-recovery MR: the effect of hybrid RARE readout on the null points of fat and cerebrospinal fluid.

نویسندگان

  • E R Melhem
  • H Jara
  • H Shakir
  • T A Gagliano
چکیده

PURPOSE To evaluate the effect of the hybrid RARE (rapid acquisition with relaxation enhancement) readout, commonly coupled to inversion-recovery pulse sequences, on the null inversiton time (TI) of fluid and fat using both phantoms and human volunteers. METHODS Two phantoms, simulating fat (phantom A) and cerebrospinal fluid (phantom B), respectively, were imaged using a fast inversion-recovery sequence that coupled an inversion-recovery preparation pulse to a hybrid RARE readout. At repetition times (TRs) ranging from 700 to 20,000, the TI necessary to null the signal from each phantom (null TI) was determined for an echo train length of 4, 6, 8, 10, 12, 14, 16, 18, and 20, respectively. Plots of null TI versus echo train length at different TRs were generated for both phantoms. Fast inversion-recovery MR imaging of the cervical spine and brain was performed in healthy volunteers. At a fixed TR and TI, the adequacy of signal suppression from bone marrow and cerebrospinal fluid was assessed as a function of echo train length. RESULTS There was a gradual decrease of null TI for both phantoms with echo train length. This decrease persisted at longer TRs for phantom B (T1 = 3175 +/- 70 milliseconds) than for phantom A (T1 = 218 +/- 5 milliseconds). In the human volunteers, there was a gradual loss of suppression of signal from bone marrow and cerebrospinal fluid, with changes in the hybrid RARE readout. CONCLUSION To optimize specific tissue suppression, radiologists implementing fast inversion-recovery MR imaging should be aware of the effects of the hybrid RARE readout on null TI.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The effect of inversion times on the minimum signal intensity of the contrast agent concentration using inversion recovery t1-weighted fast imaging sequence

  Background :Inversion recovery (IR) pulse sequences can generate T1-weighted images with a different range of inversion time (TI) to suppress or null the signal intensity (SI) for a specified tissue. In this study, we aimed to investigate the effect of TI values on the concentration of the contrast agent, which leads to a minimum signal intensity, using an inversion recovery T1-weighted 3-dim...

متن کامل

MR of the spine with a fast T1-weighted fluid-attenuated inversion recovery sequence.

PURPOSE To optimize a T1-weighted fast fluid-attenuated inversion recovery (FLAIR) sequence using computer-simulated data and to study its clinical utility for imaging the spine. METHODS Relative signal intensities and contrast of relevant normal and pathologic tissues in the spine were computed using an inversion recovery equation modified to account for a hybrid RARE (rapid acquisition with...

متن کامل

Effects of mild hypothermia therapy on the levels of glutathione in rabbit blood and cerebrospinal fluid after cardiopulmonary resuscitation

Objective(s): The aim of this study wasto investigate the effects of mild hypothermia therapy on oxidative stress injury of rabbit brain tissue after cardiopulmonary resuscitation (CPR). Materials and Methods: Rabbit models of cardiac arrest were established. After the restoration of spontaneous circulation, 50 rabbits were randomly divided into normothermia and hypothermia groups. The followi...

متن کامل

Comparison of T2-weighted and fluid-attenuated inversion-recovery fast spin-echo MR sequences in intracerebral AIDS-associated disease.

PURPOSE To compare the value of fast fluid-attenuated inversion-recovery (FLAIR) with T2-weighted fast spin-echo MR imaging in the detection of acquired immunodeficiency virus (AIDS)-related lesions of the brain. METHODS Forty-four human immunodeficiency virus (HIV)-positive patients were examined with both sequences on either a 1.0-T or a 1.5-T MR system. The number, size, location, and cons...

متن کامل

2390 Fluid attenuated inversion recovery ( FLAIR ) with readout - segmented ( rs ) - EPI

Fluid attenuated inversion recovery (FLAIR) with readout-segmented (rs)-EPI Samantha J Holdsworth, Stefan Skare, Kristen Yeom, and Michael E Moseley Lucas Center for Imaging, Department of Radiology, Stanford University, Palo Alto, CA, United States, Clinical Neuroscience, Karolinksa Institute, Stockholm, Sweden, Lucile Packard Children's Hospital, Department of Radiology, Stanford University, ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • AJNR. American journal of neuroradiology

دوره 18 9  شماره 

صفحات  -

تاریخ انتشار 1997