Leak and obstruction with mask ventilation during simulated neonatal resuscitation.

نویسندگان

  • Kim Schilleman
  • Ruben S Witlox
  • Enrico Lopriore
  • Colin J Morley
  • Frans J Walther
  • Arjan B te Pas
چکیده

OBJECTIVES To evaluate mask technique during simulated neonatal resuscitation and test the effectiveness of training in optimal mask handling. STUDY DESIGN Seventy participants(consultants, registrars and nurses) from neonatal units were asked to administer positive pressure ventilation at a flow of 8 l/min and a frequency of 40-60/min to a modified leak free, term newborn manikin (lung compliance 0.5 ml/cm H(2)O) using a Neopuff T-piece device. Recordings were made (1) before training, (2) after training in mask handling and (3) 3 weeks later. Leak was calculated. Obstruction (tidal volume <60% of optimal tidal volume) and severe obstruction (<30% of optimal tidal volume) were calculated when leak was minimal. RESULTS For the 70 participants, median (IQR) leak was 71% (32-95%) before training, 10% (5-37%) directly after training and 15% (4-33%) 3 weeks later (p<0.001). When leak was minimal, gas flow obstruction was observed before, directly after training and 3 weeks later in 46%, 42% and 37% of inflations, respectively. Severe obstruction did not occur. CONCLUSIONS Mask ventilation during simulated neonatal resuscitation was often hampered by large leaks at the face mask. Moderate airway obstruction occurred frequently when effort was taken to minimise leak. Training in mask ventilation reduced mask leak but should also focus on preventing airway obstruction.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A comparison of different mask holds for positive pressure ventilation in a neonatal manikin.

BACKGROUND Ventilation during neonatal resuscitation is typically initiated with a face mask, but may be ineffective due to leak or obstruction. OBJECTIVE To compare leak using three methods of mask hold. METHODS Medical and nursing staff regularly involved in neonatal resuscitation used the three holds (two-point, two-handed, spider) on a manikin in a random order to apply positive pressur...

متن کامل

Neonatal resuscitation 1: a model to measure inspired and expired tidal volumes and assess leakage at the face mask.

BACKGROUND Neonatal resuscitation is a common and important intervention, and adequate ventilation is the key to success. In the delivery room, positive pressure ventilation is given with manual ventilation devices using face masks. Mannequins are widely used to teach and practise this technique. During both simulated and real neonatal resuscitation, chest excursion is used to assess tidal volu...

متن کامل

Neonatal resuscitation 2: an evaluation of manual ventilation devices and face masks.

BACKGROUND The key to successful neonatal resuscitation is effective ventilation. Little evidence exists to guide clinicians in their choice of manual ventilation device or face mask. The expiratory tidal volume measured at the mask (V(TE(mask))) is a good estimate of the tidal volume delivered during simulated neonatal resuscitation. AIM To compare the efficacy of (a) the Laerdal infant resu...

متن کامل

Neonatal resuscitation 3: manometer use in a model of face mask ventilation.

BACKGROUND Adequate ventilation is the key to successful neonatal resuscitation. Positive pressure ventilation (PPV) is initiated with manual ventilation devices via face masks. These devices may be used with a manometer to measure airway pressures delivered. The expiratory tidal volume measured at the mask (V(TE(mask))) is a good estimate of the tidal volume delivered during simulated neonatal...

متن کامل

Rescue Mask with Prefabricated Leak for Reduction of Accidental Stomach Inflation During Lay Rescuer Ventilation

Objective: To evaluate the efficacy of a simple method to minimize gastric inflation in unprotected airways during resuscitation procedures. Methods: Ventilation of an airway simulator by use of a hose extended mask with leak and without leak and with or without additional oxygen (15 litres per minute). Experimental settings with artificial lung at a compliance = 300ml/ kPa and resistance = 0,5...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Archives of disease in childhood. Fetal and neonatal edition

دوره 95 6  شماره 

صفحات  -

تاریخ انتشار 2010