Performance characteristics obtained for a new 3-dimensional lutetium oxyorthosilicate-based whole-body PET/CT scanner with the National Electrical Manufacturers Association NU 2-2001 standard.

نویسندگان

  • Marco Brambilla
  • Chiara Secco
  • Marco Dominietto
  • Roberta Matheoud
  • Gianmauro Sacchetti
  • Eugenio Inglese
چکیده

UNLABELLED This article reports the results of performance measurements obtained for the lutetium oxyorthosilicate (LSO)-based whole-body PET/CT scanner Biograph 16 HI-REZ with the National Electrical Manufacturers Association (NEMA) NU 2-2001 standard. The Biograph 16 HI-REZ combines a multislice (16-slice) spiral CT scanner with a PET scanner composed of 24.336 LSO crystals arranged in 39 rings. The crystal dimensions are 4.0x4.0x20 mm3, and the crystals are organized in 13x13 blocks coupled to 4 photomultiplier tubes each. The 39 rings allow the acquisition of 81 images 2.0 mm thick, covering an axial field of view of 162 mm. The low- and high-energy thresholds are set to 425 and 650 keV, respectively, acquiring data within a 4.5-ns-wide coincidence window. METHODS Performance measurements for the LSO-based PET/CT scanner were obtained with the NEMA NU 2-2001 standard, taking into account issues deriving from the presence of intrinsic radiation. RESULTS The results obtained with the NEMA NU 2-2001 standard measurements were as follows: average transverse and axial spatial resolutions (full width at half maximum) at 1 cm and at 10 cm off axis of 4.61 (5.10) mm and 5.34 (5.91) mm, respectively; average sensitivity of 4.92 counts per second per kilobecquerel for the 2 radial positions (0 and 10 cm); 34.1% system scatter fraction; and peak noise equivalent count (NEC) rates of 84.77 kilocounts per second (kcps) at 28.73 kBq/mL (k=1 in the NEC formula; noiseless random correction) and 58.71 kcps at 21.62 kBq/mL (k=2; noisy random correction). CONCLUSION The new integrated PET/CT system Biograph 16 HI-REZ has good overall performance, with, in particular, a high resolution, a low scatter fraction, and a very good NEC response.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

PET performance measurements for an LSO-based combined PET/CT scanner using the National Electrical Manufacturers Association NU 2-2001 standard.

UNLABELLED Results of performance measurements for a lutetium oxyorthosilicate (LSO)-based PET/CT scanner using new National Electrical Manufacturers Association (NEMA) NU 2-2001 standards are reported. METHODS Performance measurements following the NU 2-2001 standards were performed on an LSO-based PET/CT scanner. In addition, issues associated with the application of the NEMA standard to LS...

متن کامل

Performance comparison of four commercial GE discovery PET/CT scanners: A monte carlo study using GATE

  Combined PET/CT scanners now play a major role in medicine for in vivo imaging in oncology, cardiology, neurology, and psychiatry. As the performance of a scanner depends not only on the scintillating material but also on the scanner design, with regards to the advent of newer scanners, there is a need to optimize acquisition protocols as well as to compare scanner ...

متن کامل

Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities.

UNLABELLED Results from a new PET/CT scanner using lutetium-yttrium oxyorthosilicate (LYSO) crystals for the PET component are presented. This scanner, which operates in a fully 3-dimensional mode, has a diameter of 90 cm and an axial field of view of 18 cm. It uses 4 x 4 x 22 mm(3) LYSO crystals arranged in a pixelated Anger-logic detector design. This scanner was designed to perform as a high...

متن کامل

Performance and limitations of positron emission tomography (PET) scanners for imaging very low activity sources.

Emerging applications for positron emission tomography (PET) may require the ability to image very low activity source distributions in the body. The performance of clinical PET scanners in the regime where activity in the field of view is <1 MBq has not previously been explored. In this study, we compared the counting rate performance of two clinical PET/CT scanners, the Siemens Biograph Revea...

متن کامل

Impact of Time-of-Flight PET/CT with a Large Axial Field of View for Reducing Whole-Body Acquisition Time.

UNLABELLED The aim of this study was to evaluate the imaging performance of 39- and 52-ring time-of-flight (TOF) PET/CT scanners. We also assessed the potential of reducing the scanning time using a 52-ring TOF PET/CT scanner. METHODS PET/CT scanners with 39- and 52-ring lutetium oxyorthosilicate detectors were evaluated. The axial fields of view were 16.2 and 21.6 cm, respectively. We used a...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of nuclear medicine : official publication, Society of Nuclear Medicine

دوره 46 12  شماره 

صفحات  -

تاریخ انتشار 2005