The Partec CyFlow Counter could provide an option for CD4+ T-cell monitoring in the context of scaling-up antiretroviral treatment at the district level in Malawi.
نویسندگان
چکیده
A study was conducted in rural Malawi to verify (a) whether the Partec CyFlow Counter((R)) for CD4+ T-cell lymphocyte counting in HIV-positive individuals could be introduced into a district hospital laboratory and (b) whether it would produce CD4 counts of acceptable quality. CD4+ cell counting was performed using the Partec CyFlow Counter and the results were compared with a reference method (FACsCount). A total of 311 blood samples were analysed and the correlation coefficient for the CyFlow Counter was 0.92 (95% CI 0.89-0.95). Mean CD4 counts using the Partec and the reference methods were 308.2 cells/microl and 316.9 cells/microl, respectively. The mean difference in CD4 count values was -8.68 cells/microl (95% CI -18.8 to 1.4). Mean intra-run variation was -6.84 cells/microl (95% CI -12.9 to 0.79). In the district laboratory setting, the instrument could accommodate up to 75 blood samples per technician per day. After being trained, local laboratory staff found the CyFlow Counter procedures simple to run and the instrument easy to manipulate. The Partec CyFlow Counter produces sufficiently reliable results and the instrument appears robust under field conditions. It could provide a new option for introducing routine CD4+ cell monitoring at the district level in the context of scaling-up antiretroviral therapy in Malawi.
منابع مشابه
Using BD Vacutainer CD4 Stabilization Tubes for Absolute Cluster of Differentiation Type 4 Cell Count Measurement on BD FacsCount and Partec Cyflow Cytometers: A Method Comparison Study from Zimbabwe
BACKGROUND Blood collected in conventional EDTA tubes requires laboratory analysis within 48 hours to provide valid CD4 cell count results. This restricts access to HIV care for patients from rural areas in resource-constraint settings due to sample transportation problems. Stabilization Tubes with extended storage duration have been developed but not yet evaluated comprehensively. OBJECTIVE ...
متن کاملReliable and Accurate CD4+ T Cell Count and Percent by the Portable Flow Cytometer CyFlow MiniPOC and “CD4 Easy Count Kit-Dry”, as Revealed by the Comparison with the Gold Standard Dual Platform Technology
BACKGROUND An accurate and affordable CD4+ T cells count is an essential tool in the fight against HIV/AIDS. Flow cytometry (FCM) is the "gold standard" for counting such cells, but this technique is expensive and requires sophisticated equipment, temperature-sensitive monoclonal antibodies (mAbs) and trained personnel. The lack of access to technical support and quality assurance programs thus...
متن کاملمدلسازی روند تغییرات سیستم ایمنی در افراد آلوده به ویروس HIV تحت درمان با داروهای ضدرتروویروسی بهوسیله مدل مارکوف
Background: After primary infection, the number of CD4 T-cells decreases with disease progress. The patient’s immunological status could inform by The CD4 T-cell counts over the time. The main purpose of this study is to assess the trend of CD4 cell count in HIV+ patient that received Antiretroviral Therapy (ART) by using a multistate Markov model to estimate transition intensities and transiti...
متن کاملLaboratory and Field Evaluation of the Partec CyFlow MiniPOC for Absolute and Relative CD4 T-Cell Enumeration
BACKGROUND A new CD4 point-of-care instrument, the CyFlow miniPOC, which provides absolute and percentage CD4 T-cells, used for screening and monitoring of HIV-infected patients in resource-limited settings, was introduced recently. We assessed the performance of this novel instrument in a reference laboratory and in a field setting in Senegal. METHODOLOGY A total of 321 blood samples were ob...
متن کاملCD4 T-Cell Enumeration in a Field Setting: Evaluation of CyFlow Counter Using the CD4 Easy Count Kit-Dry and Pima CD4 Systems
BACKGROUND Flow Cytometry (FCM) is still considered to be the method of choice for accurate CD4 enumeration. However, the use of FCM in developing countries is problematic due to their cost and complexity. Lower-cost technologies have been introduced. We evaluated CyFlow Counter together with its lyophilized reagents, and Pima CD4 in high-temperature area, using FACSCount as reference. MATERI...