Dialysate flow rate and delivered Kt/Vurea for dialyzers with enhanced dialysate flow distribution.

نویسندگان

  • Richard A Ward
  • John W Idoux
  • Hiba Hamdan
  • Rosemary Ouseph
  • Thomas A Depner
  • Thomas A Golper
چکیده

BACKGROUND AND OBJECTIVES Previous in vitro and clinical studies showed that the urea mass transfer-area coefficient (K(o)A) increased with increasing dialysate flow rate. This observation led to increased dialysate flow rates in an attempt to maximize the delivered dose of dialysis (Kt/V(urea)). Recently, we showed that urea K(o)A was independent of dialysate flow rate in the range 500 to 800 ml/min for dialyzers incorporating features to enhance dialysate flow distribution, suggesting that increasing the dialysate flow rate with such dialyzers would not significantly increase delivered Kt/V(urea). DESIGN, SETTING, PARTICIPANTS, & MEASUREMENTS We performed a multi-center randomized clinical trial to compare delivered Kt/V(urea) at dialysate flow rates of 600 and 800 ml/min in 42 patients. All other aspects of the dialysis prescription, including treatment time, blood flow rate, and dialyzer, were kept constant for a given patient. Delivered single-pool and equilibrated Kt/V(urea) were calculated from pre- and postdialysis plasma urea concentrations, and ionic Kt/V was determined from serial measurements of ionic dialysance made throughout each treatment. RESULTS Delivered Kt/V(urea) differed between centers; however, the difference in Kt/V(urea) between dialysate flow rates of 800 and 600 ml/min was NS by any measure (95% confidence intervals of -0.064 to 0.024 for single-pool Kt/V(urea), -0.051 to 0.023 for equilibrated Kt/V(urea), and -0.029 to 0.099 for ionic Kt/V). CONCLUSIONS These data suggest that increasing the dialysate flow rate beyond 600 ml/min for these dialyzers offers no benefit in terms of delivered Kt/V(urea).

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The patient as a limit to dialysis technology.

Ward et al. show in this issue of CJASN that Kt/V urea does not improve by increasing the dialysate flow rate in dialyzers in which dialysate flow distribution has been optimized by changes in design (1). Such improved dialyzer designs were developed during the last decade, based on the awareness that poor dialysate flow distribution and the occurrence of preferential dialysate flow paths negat...

متن کامل

Effect of increasing dialysate flow rate on diffusive mass transfer of urea, phosphate and β2-microglobulin during clinical haemodialysis

BACKGROUND Diffusive clearance depends on blood and dialysate flow rates and the overall mass transfer area coefficient (K(o)A) of the dialyzer. Although K(o)A should be constant for a given dialyzer, urea K(o)A has been reported to vary with dialysate flow rate possibly because of improvements in flow distribution. This study examined the dependence of K(o)A for urea, phosphate and β(2)-microg...

متن کامل

Effect of Low Dialysate Flow Rate on Hemodialyzer Mass Transfer Area Coefficients for Urea and Creatinine.

Recent work has shown that the dialyzer mass transfer area coefficient (Ko A) for urea increases when the dialysate flow rate is increased from 500 to 800 mL/min. In this study we determined urea and creatinine clearances for two commercial dialyzers containing polysulfone hollow fibers in vitro at 37°C, a nominal blood flow rate of 300 mL/ min, and dialysate flow rates (Qd ) ranging from 100 t...

متن کامل

Removal of uraemic retention solutes in standard bicarbonate haemodialysis and long-hour slow-flow bicarbonate haemodialysis.

BACKGROUND Several studies already stressed the importance of haemodialysis (HD) time in the removal of uraemic toxins. In those studies, however, also the amount of dialysate and/or processed blood was altered. The present study aimed to investigate the isolated effect of the factor time t (by processing the same total blood and dialysate volume in two different time schedules) on the removal ...

متن کامل

Computational flow modeling in hollow-fiber dialyzers.

A three-dimensional finite volume model of the blood-dialysate interface over the complete length of the dialyzer was developed. Different equations govern dialyzer flow and pressure distribution (Navier-Stokes) and radial transport (Darcy). Blood was modeled as a non-Newtonian fluid with a viscosity varying in radial and axial direction determined by the local hematocrit, the diameter of the c...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical journal of the American Society of Nephrology : CJASN

دوره 6 9  شماره 

صفحات  -

تاریخ انتشار 2011