Ppb-level formaldehyde detection using a CW room-temperature interband cascade laser and a miniature dense pattern multipass gas cell.
نویسندگان
چکیده
A ppb-level formaldehyde (H2CO) sensor was developed using a thermoelectrically cooled (TEC), continuous-wave (CW) room temperature interband cascade laser (ICL) emitting at 3.59 μm and a miniature dense pattern multipass gas cell with >50 m optical path length. Performance of the sensor was investigated with two measurement schemes: direct absorption (DAS) and wavelength modulation spectroscopy (WMS). With an integration time of less than 1.5 second, a detection limit of ~3 ppbv for H2CO measurement with precision of 1.25 ppbv for DAS and 0.58 ppbv for WMS, respectively, was achieved without zero air based background subtraction. An Allan-Werle variance analysis indicated that the precisions can be further improved to 0.26 ppbv @ 300s for DAS and 69 pptv @ 90 s for WMS, respectively. A side-by-side comparison between two measurement schemes is also discussed in detail.
منابع مشابه
Interband cascade laser based absorption sensor for ppb-level formaldehyde detection
A trace gas absorption sensor for formaldehyde (H2CO) detection was developed using a continuous wave, room temperature, low-power consumption interband cascade laser (ICL) at 3.6 μm. The recent availability of ICLs with wavelength ranged between 34 μm enables the sensitive detection of trace gases such as formaldehyde that possesses a strong absorption band in this particular wavelength regio...
متن کاملInfrared Dual - Gas CH 4 / C 2 H 6 Sensor Using Two Continuous - Wave Interband Cascade Lasers
Abstract—An infrared dual-gas sensor system for the simultaneous detection and monitoring of methane (CH4) and ethane (C2H6) at parts-per-billion by volume (ppbv) concentration levels was developed using two room temperature, distributed feedback (DFB) interband cascade lasers (ICLs) and two miniature multipass cells with an effective absorption length of 54.6 m. Laser direct absorption spect...
متن کاملCW EC-QCL-based sensor for simultaneous detection of H2O, HDO, N2O and CH4 using multi-pass absorption spectroscopy
A sensor system based on a continuous wave, external-cavity quantum-cascade laser (CW EC-QCL) was demonstrated for simultaneous detection of atmospheric H2O, HDO, N2O and CH4 using a compact, dense pattern multi-pass gas cell with an effective path-length of 57.6 m. The ECQCL with a mode-hop-free spectral range of 1225-1285 cm operating at ~7.8 μm was scanned covering four neighboring absorptio...
متن کاملEffective utilization of quantum-cascade distributed-feedback lasers in absorption spectroscopy.
A variable duty cycle quasi-cw frequency scanning technique was applied to reduce thermal effects resulting from the high heat dissipation of type I quantum-cascade lasers. This technique was combined with a 100-m path-length multipass cell and a zero-air background-subtraction technique to enhance detection sensitivity to a parts-in-10(9) (ppb) concentration level for spectroscopic trace-gas d...
متن کاملSimultaneous atmospheric nitrous oxide, methane and water vapor detection with a single continuous wave quantum cascade laser.
A continuous wave (CW) quantum cascade laser (QCL) based absorption sensor system was demonstrated and developed for simultaneous detection of atmospheric nitrous oxide (N(2)O), methane (CH(4)), and water vapor (H(2)O). A 7.73-µm CW QCL with its wavelength scanned over a spectral range of 1296.9-1297.6 cm(-1) was used to simultaneously target three neighboring strong absorption lines, N(2)O at ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Optics express
دوره 23 15 شماره
صفحات -
تاریخ انتشار 2015