Flow cytometric reticulocyte counting. Parallel evaluation of five fully automated analyzers: an NCCLS-ICSH approach.

نویسندگان

  • M Buttarello
  • P Bulian
  • G Farina
  • V Temporin
  • L Toffolo
  • E Trabuio
  • P Rizzotti
چکیده

We performed a parallel evaluation of 5 automated reticulocyte analyzers. The guidelines were those proposed by the National Committee for Clinical Laboratory Standards and the International Council for Standardisation in Haematology. Duplicate analyses were performed for 225 healthy subjects and 115 patients affected by various diseases. The reference intervals were different for each method (ADVIA 120, 27-125 x 10(3)/microL [27-125 x 10(9)/L]; CELL DYN 4000, 25-108 x 10(3)/microL [25-108 x 10(9)/L]; GEN-S, 20-85 x 10(3)/microL [20-85 x 10(9)/L]; SE 9500 RET, 23-95 x 10(3)/microL [23-95 x 10(9)/L]; and VEGA RETIC, 30-130 x 10(3)/microL [30-130 x 10(9)/L]). The comparisons of percentage counts with the microscopic reference method were satisfactory for all automated methods. However, a tendency to overestimate at low counts was noted. This progressively increased from the SE 9500 RET to the VEGA RETIC. The imprecision was excellent for all the methods at normal and high concentrations. This was higher at low concentrations. When compared with the microscopic reference, the analyzers showed satisfactory sensitivity at low counts and excellent sensitivity at high counts. The overall agreement varied from 74.8% for the GEN-S to 86.1% for the SE 9500 RET.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

ICSH guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting.

This revision is intended to update the 1994 ICSH guidelines. It is based on those guidelines but is updated to include new methods, such as digital image analysis for blood cells, a flow cytometric method intended to replace the reference manual 400 cell differential, and numerous new cell indices not identified morphologically are introduced. Haematology analysers are becoming increasingly co...

متن کامل

Comment on: Evaluation of erythrocyte and reticulocyte parameters as indicative of iron deficiency in patients with anemia of chronic disease☆

The article on the evaluation of the effectiveness of mature red cell and reticulocyte parameters under three conditions: iron deficiency anemia, anemia of chronic disease (ACD), and anemia of chronic disease associated with absolute iron deficiency by Torino et al.1 is very valuable. Automated reticulocyte counts are widely used in the clinical laboratory due to their greater precision, accura...

متن کامل

Reticulocyte counting using flow cytometry.

A flow cytometric method for the quantitation of reticulocytes was refined for routine laboratory use. Blood (2 microliters) is added to 2 ml of 0.4 microM thiazole orange in phosphate buffered saline, incubated at room temperature for 90 minutes, and analysed on a Coulter EPICS Profile flow cytometer, with gating for red cells on the basis of forward and right angled light scatter. Blood (2 mi...

متن کامل

Multivariate statistical modeling: alternative approach to test evaluation, applied to counting reticulocytes by flow cytometry.

We present a statistical path analysis model for the evaluation of two tests in the absence of a "gold standard" method. This model is applied to the evaluation of flow-cytometric and visual reticulocyte counting by using as the comparison method a combination of three hematological measurements: hemoglobin concentration (HGB), mean cellular volume (MCV), and erythrocyte density width (EDW). We...

متن کامل

Development and Evaluation of a Novel In-Clinic Automated Hematology Analyzer, ProCyte Dx, for Canine Erythrocyte Indices, Leukogram, Platelet Counts and Reticulocyte Counts

A novel hematology analyzer for small animal medicine, ProCyte Dx, was developed from combination of the fluorescence laser flow cytometry and laminar flow impedance technologies, and its accuracy was evaluated by comparing with the conventional impedance-based hematology analyzer, pocH-100iV Diff, or microscopic manual cell counting methods with staining blood smears in the canine blood. Blood...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • American journal of clinical pathology

دوره 115 1  شماره 

صفحات  -

تاریخ انتشار 2001