Superiority of automatic remote monitoring compared with in-person evaluation for scheduled ICD follow-up in the TRUST trial - testing execution of the recommendations

نویسندگان

  • Niraj Varma
  • Justin Michalski
  • Bruce Stambler
  • Behzad B. Pavri
چکیده

AIMS To test recommended implantable cardioverter defibrillator (ICD) follow-up methods by 'in-person evaluations' (IPE) vs. 'remote Home Monitoring' (HM). METHODS AND RESULTS ICD patients were randomized 2:1 to automatic HM or to Conventional monitoring, with follow-up checks scheduled at 3, 6, 9, 12, and 15 months post-implant. Conventional patients were evaluated with IPE only. Home Monitoring patients were assessed remotely only for 1 year between 3 and 15 month evaluations. Adherence to follow-up was measured. HM and Conventional patients were similar (age 63 years, 72% male, left ventricular ejection fraction 29%, primary prevention 73%, DDD 57%). Conventional management suffered greater patient attrition during the trial (20.1 vs. 14.2% in HM, P = 0.007). Three month follow-up occurred in 84% in both groups. There was 100% adherence (5 of 5 checks) in 47.3% Conventional vs. 59.7% HM (P < 0.001). Between 3 and 15 months, HM exhibited superior (2.2×) adherence to scheduled follow-up [incidence of failed follow up was 146 of 2421 (6.0%) in HM vs. 145 of 1098 (13.2%) in Conventional, P < 0.001] and punctuality. In HM (daily transmission success rate median 91%), transmission loss caused only 22 of 2275 (0.97%) failed HM evaluations between 3 and 15 months; others resulted from clinic oversight. Overall IPE failure rate in Conventional [193 of 1841 (10.5%) exceeded that in HM [97 of 1484 (6.5%), P < 0.001] by 62%, i.e. HM patients remained more loyal to IPE when this was mandated. CONCLUSION Automatic remote monitoring better preserves patient retention and adherence to scheduled follow-up compared with IPE. CLINICAL TRIAL REGISTRATION NCT00336284.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Automatic remote monitoring of implantable cardioverter-defibrillator lead and generator performance: the Lumos-T Safely RedUceS RouTine Office Device Follow-Up (TRUST) trial.

BACKGROUND Monitoring performance of implantable cardioverter-defibrillator (ICD) generators and leads is important. Methods available are with in-person evaluations or by automatic remote home monitoring (HM). These were prospectively evaluated and compared in the TRUST trial. The HM technology tested performed daily self-checks and databasing with rapid event notifications for out-of-range (i...

متن کامل

The Lumos-T Safely RedUceS RouTine Office Device Follow-Up (TRUST) Trial

Background—Monitoring performance of implantable cardioverter-defibrillator (ICD) generators and leads is important. Methods available are with in-person evaluations or by automatic remote home monitoring (HM). These were prospectively evaluated and compared in the TRUST trial. The HM technology tested performed daily self-checks and databasing with rapid event notifications for out-of-range (i...

متن کامل

Quarterly vs. yearly clinical follow-up of remotely monitored recipients of prophylactic implantable cardioverter-defibrillators: results of the REFORM trial

AIMS The rapidly increasing number of patients with implantable cardioverter-defibrillators (ICD) places a large burden on follow-up providers. This study investigated the possibility of longer in-office follow-up intervals in primary prevention ICD patients under remote monitoring with automatic daily data transmissions from the implant memory. METHODS AND RESULTS Conducted in 155 ICD recipi...

متن کامل

Percutaneous repair of sinus venosus defect with anomalous pulmonary venous return.

pace 2013;15(Suppl 1):i14–i16. 8. Varma N, Epstein A, Irimpen A, Schweikert R, Shah J, Love CJ. Efficacy and safety of automatic remote monitoring for ICD follow-up: the TRUST trial. Circulation 2010; 122:325–332. 9. Varma N. Rationale and design of a prospective study of the efficacy of a remote monitoring system used in ICD follow-up: the Lumos-T reduces routine office device follow-up study ...

متن کامل

Arrhythmia/Electrophysiology Efficacy and Safety of Automatic Remote Monitoring for Implantable Cardioverter-Defibrillator Follow-Up The Lumos-T Safely Reduces Routine Office Device Follow-Up (TRUST) Trial

Background—Monitoring implantable cardiac device function and patient condition is important. The Lumos-T Safely Reduces Routine Office Device Follow-Up (TRUST) trial tested the hypothesis that remote home monitoring with automatic daily surveillance (HM) is safe and effective for implantable cardioverter-defibrillator follow-up for 1 year and enables rapid physician evaluation of significant e...