Efficiency calibration of ITER bolometer prototypes Broad-band efficiency calibration of ITER bolometer prototypes using Pt absorbers on SiN membranes
نویسندگان
چکیده
Broad-band efficiency calibration of ITER bolometer prototypes using Pt absorbers on SiN membranes H Meister, M Willmeroth, a) D Zhang, and , A Gottwald, M Krumrey, F Scholze Max-Planck-Institut für Plasmaphysik (IPP), EURATOM Association, Boltzmannstr. 2, 85748 Garching, Germany Max-Planck-Institut für Plasmaphysik (IPP), EURATOM Association, Teilinstitut Greifswald, Wendelsteinstraße 1, 17491 Greifswald, Germany Physikalisch-Technische Bundesanstalt (PTB), Abbestraße 2-12, 10587 Berlin, Germany
منابع مشابه
Broad-band efficiency calibration of ITER bolometer prototypes using Pt absorbers on SiN membranes.
The energy resolved efficiency of two bolometer detector prototypes for ITER with 4 channels each and absorber thicknesses of 4.5 μm and 12.5 μm, respectively, has been calibrated in a broad spectral range from 1.46 eV up to 25 keV. The calibration in the energy range above 3 eV was performed against previously calibrated silicon photodiodes using monochromatized synchrotron radiation provided ...
متن کاملReliability issues for a bolometer detector for ITER at high operating temperatures.
The first detector prototypes for the ITER bolometer diagnostic featuring a 12.5 μm thick Pt-absorber have been realized and characterized in laboratory tests. The results show linear dependencies of the calibration parameters and are in line with measurements of prototypes with thinner absorbers. However, thermal cycling tests up to 450 °C of the prototypes with thick absorbers demonstrated th...
متن کاملOptimization of a bolometer detector for ITER based on Pt absorber on SiN membrane.
Any plasma diagnostic in ITER must be able to operate at temperatures in excess of 200 °C and neutron loads corresponding to 0.1 dpa over its lifetime. To achieve this aim for the bolometer diagnostic, a miniaturized metal resistor bolometer detector based on Pt absorbers galvanically deposited on SiN membranes is being developed. The first two generations of detectors featured up to 4.5 μm thi...
متن کاملThe ITER bolometer diagnostic: status and plans.
A consortium consisting of four EURATOM Associations has been set up to develop the project plan for the full development of the ITER bolometer diagnostic and to continue urgent R&D activities. An overview of the current status is given, including detector development, line-of-sight optimization, performance analysis as well as the design of the diagnostic components and their integration in IT...
متن کاملSilicon nitride Micromesh Bolometer Array for Submillimeter Astrophysics.
We present the design and performance of a feedhorn-coupled bolometer array intended for a sensitive 350-mum photometer camera. Silicon nitride micromesh absorbers minimize the suspended mass and heat capacity of the bolometers. The temperature transducers, neutron-transmutation-doped Ge thermistors, are attached to the absorber with In bump bonds. Vapor-deposited electrical leads address the t...