MRI-Guided Stereotactic System for Delivery of Intraspinal Microstimulation.
نویسندگان
چکیده
STUDY DESIGN Laboratory/animal-based proof of principle study. OBJECTIVE To validate the accuracy of a magnetic resonance imaging (MRI)-guided stereotactic system for intraspinal electrode targeting and demonstrate the feasibility of such a system for controlling implantation of intraspinal electrodes. SUMMARY OF BACKGROUND DATA Intraspinal microstimulation (ISMS) is an emerging preclinical therapy, which has shown promise for the restoration of motor function following spinal cord injury. However, targeting inaccuracy associated with existing electrode implantation techniques remains a major barrier preventing clinical translation of ISMS. METHODS System accuracy was evaluated using a test phantom comprised of nine target locations. Targeting accuracy was determined by calculating the root mean square error between MRI-generated coordinates and actual frame coordinates required to reach the target positions. System performance was further validated in an anesthetized pig model by performing MRI-guided intraspinal electrode implantation and stimulation followed by computed tomography of electrode location. Finally, system compatibility with a commercially available microelectrode array was demonstrated by implanting the array and applying a selection of stimulation amplitudes that evoked hind limb responses. RESULTS The root mean square error between actual frame coordinates and software coordinates, both acquired using the test phantom, was 1.09 ± 0.20 mm. Postoperative computed tomography in the anesthetized pig confirmed spatially accurate electrode placement relative to preoperative MRI. Additionally, MRI-guided delivery of a microwire electrode followed by ISMS evoked repeatable electromyography responses in the biceps femoris muscle. Finally, delivery of a microelectrode array produced repeatable and graded hind limb evoked movements. CONCLUSION We present a novel frame-based stereotactic system for targeting and delivery of intraspinal instrumentation. This system utilizes MRI guidance to account for variations in anatomy between subjects, thereby improving upon existing ISMS electrode implantation techniques. LEVEL OF EVIDENCE N/A.
منابع مشابه
Restoring Motor Functions in Paralyzed Limbs through Intraspinal Multielectrode Microstimulation Using Fuzzy Logic Control and Lag Compensator
In this paper, a control strategy is proposed for control of ankle movement on animals using intraspinal microstimulation (ISMS). The proposed method is based on fuzzy logic control. Fuzzy logic control is a methodology of intelligent control that mimics human decision-making process. This type of control method can be very useful for the complex uncertain systems that their mathematical model ...
متن کاملA robust MRI-compatible system to facilitate highly accurate stereotactic administration of therapeutic agents to targets within the brain of a large animal model
Achieving accurate intracranial electrode or catheter placement is critical in clinical practice in order to maximise the efficacy of deep brain stimulation and drug delivery respectively as well as to minimise side-effects. We have developed a highly accurate and robust method for MRI-guided, stereotactic delivery of catheters and electrodes to deep target structures in the brain of pigs. This...
متن کاملCyberKnife radiosurgery for spinal neoplasms.
The role of stereotactic radiosurgery for the treatment of intracranial lesions is well established. Its use for the treatment of spinal lesions has been limited by the availability of effective target immobilization and localization technologies. Conventional external beam radiotherapy lacks the precision to allow delivery of large doses of radiation near radiosensitive structures such as the ...
متن کاملEvaluation of Thiel cadaveric model for MRI-guided stereotactic procedures in neurosurgery
BACKGROUND Magnetic resonance imaging (MRI)-guided deep brain stimulation (DBS) and high frequency focused ultrasound (FUS) is an emerging modality to treat several neurological disorders of the brain. Developing reliable models to train and assess future neurosurgeons is paramount to ensure safety and adequate training of neurosurgeons of the future. METHODS We evaluated the use of Thiel cad...
متن کاملNovel platform for MRI-guided convection-enhanced delivery of therapeutics: preclinical validation in nonhuman primate brain.
BACKGROUND/AIMS A skull-mounted aiming device and integrated software platform has been developed for MRI-guided neurological interventions. In anticipation of upcoming gene therapy clinical trials, we adapted this device for real-time convection-enhanced delivery of therapeutics via a custom-designed infusion cannula. The targeting accuracy of this delivery system and the performance of the in...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Spine
دوره 41 13 شماره
صفحات -
تاریخ انتشار 2016