Design and implementation of a multifrequency near-infrared diffuse optical tomography system.
نویسندگان
چکیده
The design and implementation of a multifrequency and multispectral diffuse optical tomography system is described. Four wavelengths are utilized: 665, 785, 808, and 830 nm. The system is based on a network analyzer, which provides rf modulation signals for the laser diodes, as well as measures the amplitude and the phase of the detected signals. Six different modulation frequencies ranging from 110 to 280 MHz are used. The details of instrumentation, calibration, data acquisition, and performance of the system are given. A finite element algorithm is used to solve the diffusion equation, and an inverse solver based on this forward solver is implemented to calculate the absorption and scattering maps from the acquired data. Data acquisition for one wavelength is completed in less than 2.5 min for a single modulation frequency. The measurement repeatability is 0.5% in ac intensity and 0.2 deg in phase. The performance of the system is evaluated with phantom studies. A multifrequency reconstruction algorithm is used, in which a single absorption and scattering image pair is obtained using the whole dataset obtained at different modulation frequencies. It is shown that the multifrequency reconstruction approach provides superior image quality compared to the single frequency counterpart.
منابع مشابه
Validation of Hemoglobin and Water Molar Absorption Spectra in Near-Infrared Diffuse Optical Tomography
Near Infrared diffuse optical tomography was used to measure the molar absorption spectra for deoxyhemoglobin, oxyhemoglobin and water through phantom experiments. The measured spectra compensate for any systemic errors and errors arising from model-based estimation of concentrations of these chromophores, and show reasonable agreement with existing literature values. The spectra were validated...
متن کاملA Generalized Least-squares minimization method for near infrared diffuse optical tomography
Diffuse optical tomography (DOT) has the potential to become a non-invasive and non-ionizing diagnostic imaging technique for breast cancer imaging. DOT uses near-infrared (NIR) light to illuminate the breast, generally through the use of fiber optic bundles, and the diffusely transmitted light is collected from the tissue surface. To get functional images of the breast, one needs to use advanc...
متن کاملInstrumentation for video-rate near-infrared diffuse optical tomography
This article describes the design, rationale, and system performance of a rapid imaging near-infrared diffuse optical tomography system that is capable of collecting tomographic measurements at video rate. Data-acquisition speed of 35 frames/s is achieved by spectral encoding of the sources, followed by spectral decoding of all detection channels in parallel in a spectrometer and using charge-c...
متن کاملA Near-Infrared Optical Tomography System Based on Photomultiplier Tube
Diffuse optical tomography (DOT) is a rapidly growing discipline in recent years. It plays an important role in many fields, such as detecting breast cancer and monitoring the cerebra oxygenation. In this paper, a relatively simple, inexpensive, and conveniently used DOT system is presented in detail, in which only one photomultiplier tube is employed as the detector and an optical multiplexer ...
متن کاملImage quality improvement of diffuse optical tomography of breast tumor using artificial intelligence
This article has no abstract.
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of biomedical optics
دوره 11 1 شماره
صفحات -
تاریخ انتشار 2006