Accuracy of a manual torque application device for morse-taper implants: a technical note.
نویسندگان
چکیده
PURPOSE The objective of this study was to compare torques applied by new and used manual torque devices for Morse-taper implants. MATERIAL AND METHODS Fifteen ITI manual torque devices were tested. Those in group 1 (n = 5) were new (ie, never used), those in group 2 (n = 5) had been used 50 to 200 times, and those in group 3 (n = 5) had been used 500 to 1,000 times. The torques applied by each device were measured for 35 Ncm and 15 Ncm targets in an experimental setup by a custom-made wrench with strain gauges connected to a data acquisition system. The strain-gauge signals were simultaneously delivered to a computer at a sample rate of 10,000 Hz and converted to torque units. RESULTS New devices applied higher torques than used devices for the 35-Ncm torque target (P < .05). The torques applied by group 3 devices were approximately 1.5 Ncm lower than those of other groups for the 35-Ncm target and approximately 1 Ncm lower for the 15-Ncm target. DISCUSSION AND CONCLUSION ITI manual torque devices deliver consistent torque output, although a slight decrease occurs as a consequence of clinical use.
منابع مشابه
In vitro microbiological analysis of bacterial seal at the implant-abutment interface using two morse taper implant models.
The objective of this study was to evaluate the bacterial seal at the implant-abutment interface using two morse taper implant models, by means of an in vitro microbiological analysis. For that were used 15 implants with mini-abutments tightened by friction, no screws (Group 1); and 30 implants with screw-tightened abutments, of which 15 received 20 N.cm of closing torque (Group 2) and the othe...
متن کاملAn Indirect Method to Measure Abutment Screw Preload: A Pilot Study Based on Micro-CT Scanning.
This study aimed to measure the preload in different implant platform geometries based on micro-CT images. External hexagon (EH) implants and Morse Tapered (MT) implants (n=5) were used for the preload measurement. The abutment screws were scanned in micro-CT to obtain their virtual models, which were used to record their initial length. The abutments were screwed on the implant with a 20 Ncm t...
متن کاملRemoval torques of conical, tapered implant abutments: the effects of anodization and reduction of surface area.
PURPOSE To examine the effects of anodization (surface coating) and reduction of internal Morse taper surface area on the reversal torque values of Straumann ITI dental implants and abutments. MATERIALS AND METHODS Eighty ITI solid screw implants were mated with corresponding 5.5-mm solid abutments. The assemblies were divided into 4 test groups of 20 specimens. All abutments were torque tigh...
متن کاملEvaluation of the ITI Morse taper implant/abutment design with an internal modification.
PURPOSE The purpose of this study was to compare internal Morse taper connections in 2 separate modes: repeated torque/reverse-torque values and compressive bending at a 30-degree off-axis angle. MATERIALS AND METHODS Three sample groups (n = 12 in each group)--a solid-screw implant paired with a 5.5-mm solid abutment (SSI), a synOcta implant with a 5.5-mm solid abutment (SOI), and a synOcta ...
متن کاملPreload loss of abutment screws after dynamic fatigue in single implant-supported restorations. A systematic review
Background To carry out a systematic literature review of the causes of preload loss of the abutment screws, of internal and external connection implants, tightened to different torque values and subjected to cyclic loading. Material and Methods A systematic search was conducted in PubMed, EMBASE, and Cochrane Library databases with reference to in vitro studies in which internal and external...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The International journal of oral & maxillofacial implants
دوره 19 5 شماره
صفحات -
تاریخ انتشار 2004