Effects of the transcutaneous electrode temperature on the accuracy of transcutaneous carbon dioxide tension
نویسندگان
چکیده
AIM The harmful effect of hypocapnia on the neonatal brain emphasizes the importance of monitoring arterial carbon dioxide tension (PaCO2). Transcutaneous monitoring of carbon dioxide (tcPCO2) reduces the need for arterial blood sampling. Drawbacks are high electrode temperature causing risks of skin burning. The aim was to determine the accuracy and precision of tcPCO2 at reduced electrode temperature. METHODS Forty newborns (GA 24.9-41.7) were included. Two tc-monitors were applied (TCM4, Radiometer, Copenhagen). Arterial blood gas sampling and monitoring of tcPCO2-level at different electrode temperatures was done simultaneously (39°C, 40°C, 41°C, 42°C, 44°C). Difference of PaCO2 - tcPCO2 was expressed as a percentage of the mean. RESULTS Mean PaCO2 was 5.8kPa [3,2; 7.9]. Bias (PaCO2 - tcPCO2) increased from 5% at 44°C to 17% at 39°C, but did not differ significantly between 41°C and 40°C. The precision of the tcPCO2 at each temperature ranged from +7-10%. After correction for the temperature-dependent overreading, we found increasing PaCO2 - tcPCO2 difference with increasing PaCO2, approx. 2% pr. kPa increase of CO(2). Only mild transient erythema was observed. CONCLUSION A lower electrode temperature in tcPCO2-monitoring increases systematic overreading of the tc-electrode. However, in very preterm babies, monitoring at 40°C or 41°C is possible provided a bias correction of 12-15% is applied.
منابع مشابه
Transcutaneous and capillary pCO2 and pO2 measurements in healthy adults.
Transcutaneous carbon dioxide and oxygen tensions (tc-pCO2 and tc-pO2) were measured in seven healthy adult volunteers during hyperventilation in atmospheric air and during CO2 inhalation. Three skin sensors were applied to each subject: an O2 electrode, a CO2 electrode, and a combined O2-CO2 electrode, each heated to 44 degrees C. We observed close correlation between tc-pCO2 and capillary-pCO...
متن کاملEffects of theophylline on ventilatory response to hypoxic challenge.
The respiratory and arousal responses to mild hypoxia during quiet sleep were studied using inductive plethysmography and transcutaneous gas electrodes in 11 apnoeic infants before and after the administration of oral theophylline (3 mg/kg). Theophylline changed the ventilatory response to a more biphasic pattern--that is, ventilation decreased after an initial increase. The relative ventilator...
متن کاملTranscutaneous estimation of arterial carbon dioxide in intensive care. Which electrode temperature?
Transcutaneous and arterial carbon dioxide were measured simultaneously in 57 children (age range 10 days to 14.3 years) undergoing intensive care. All were haemodynamically stable at the time of study. Mean calibration time with 5 and 10% carbon dioxide was 43 (range 38-58) minutes and mean arterialisation time was 10.5 (range 3-30) minutes. Duplicate hourly arterial samples over a four hour p...
متن کاملUse of a combined oxygen and carbon dioxide transcutaneous electrode in the estimation of gas exchange during exercise.
BACKGROUND Accurate and reliable measurement of gas exchange during exercise has traditionally involved arterial cannulation. Non-invasive devices to estimate arterial oxygen (O2) and carbon dioxide (CO2) tensions are now available. A method has been devised and evaluated for measuring gas exchange during exercise with a combined transcutaneous O2 and CO2 electrode. METHODS Symptom limited ex...
متن کاملValidity of transcutaneous oxygen/carbon dioxide pressure measurement in the monitoring of mechanical ventilation in stable chronic respiratory failure.
The accuracy and precision of transcutaneous pressure measurements of oxygen (Ptc,O2) and carbon dioxide (Ptc,CO2) in the monitoring of nocturnal assisted ventilation in adult patients were evaluated. Transcutaneous measurements obtained with two analysers, Radiometer TINATCM3 (R) and Kontron MicroGas-7650 (K), were compared with arterial blood gases analysed in blood samples withdrawn simultan...