Eye-Tracker-Guided Non-Mechanical Excimer Laser Assisted Penetrating Keratoplasty
نویسندگان
چکیده
PURPOSE The purpose of the study was to implement a new eye tracking mask which could be used to guide the laser beam in automated non-mechanical excimer laser assisted penetrating keratoplasty. MATERIALS AND METHODS A new trephination mask design with an elevated surface geometry has been proposed with a step formation between conical and flat interfaces. Two recipient masks of 7.5/8.0 mm have been manufactured and tested. The masks have outer diameter of 12.5 mm, step formation at 10.5 mm, and slope of conical surfaces 15°. Its functionality has been tested in different lateral positions and tilts on a planar surface, and pig eye experiments. After successful validation on porcine eyes, new masks have been produced and tested on two patients. RESULTS The build-in eye tracking software of the MEL 70 was always able to capture the masks. It has been shown that the unwanted pigmentation/pattern induced by the laser pulses on the mask surface does not influence the eye-tracking efficiency. The masks could be tracked within the 18 × 14 mm lateral displacement and up to 12° tilt. Two patient cases are demonstrated. No complications were observed during the surgery, although it needs some attention for aligning the mask horizontally before trephination. Stability of eye tracking masks is emphasized by inducing on purpose movements of the patient head. CONCLUSION Eye-tracking-guided penetrating keratoplasty was successfully applied in clinical practice, which enables robust tracking criteria within an extended range. It facilitates the automated trephination procedure of excimer laser-assisted penetrating keratoplasty.
منابع مشابه
Optical pachymetry-guided custom excimer laser-assisted lamellar keratoplasty for the surgical treatment of keratoconus.
PURPOSE To evaluate the anatomic and functional results of optical pachymetry-guided custom excimer laser-assisted lamellar keratoplasty in keratoconus patients. SETTING Eye Clinic, University of L'Aquila, L'Aquila, Italy. DESIGN Prospective noncomparative case series. METHODS Patients with keratoconus having unilateral surgery using custom excimer laser-assisted lamellar keratoplasty wer...
متن کاملTopography-guided (NIDEK customized aspheric treatment zone) photorefractive keratectomy with mitomycin C after penetrating keratoplasty for keratoconus: case report.
PURPOSE To report topography-guided photorefractive keratectomy (PRK) with mitomycin C (MMC) after penetrating keratoplasty for keratoconus. METHODS A 34-year-old woman with irregular astigmatism after penetrating keratoplasty in the right eye underwent PRK. Topography-guided surface ablation using the customized aspheric treatment zone ablation (CATz) was programmed for a 5.00-mm optical zon...
متن کاملExcimer versus Femtosecond Laser Assisted Penetrating Keratoplasty in Keratoconus and Fuchs Dystrophy: Intraoperative Pitfalls
Purpose. To assess the intraoperative results comparing two non-mechanical laser assisted penetrating keratoplasty approaches in keratoconus and Fuchs dystrophy. Patients and Methods. 68 patients (age 18 to 87 years) with keratoconus or Fuchs dystrophy were randomly distributed to 4 groups. 35 eyes with keratoconus and 33 eyes with Fuchs dystrophy were treated with either excimer laser ([Exc] g...
متن کاملExcimer laser-assisted lamellar keratoplasty for the surgical treatment of keratoconus.
PURPOSE To evaluate the anatomical and functional results of excimer laser-assisted lamellar keratoplasty (ELLK) in keratoconus patients. SETTING Eye Clinic, University of L'Aquila, L'Aquila, Italy. METHODS This prospective case series comprised patients with keratoconus who had ELLK and were examined preoperatively and 3, 6, 12, and 24 months postoperatively. Outcome measures were uncorrec...
متن کامل"Orientation teeth" in non-mechanical laser corneal trephination for penetrating keratoplasty: 2.94 microm Er:YAG v 193 nm ArF excimer laser.
BACKGROUND/AIMS "Orientation teeth" at the donor trephination margin and correspondent "notches" at the host margin facilitate graft orientation and avoid "horizontal torsion" induced by asymmetric suture placement. In this study the quality and reproducibility of these structures created by non-mechanical laser corneal trephination were compared using two laser emissions. METHODS The procedu...
متن کامل