Sentinel node detection failure due to defective labeling and large particle size of Tc - 99 m antimony sulfide colloid
نویسندگان
چکیده
Introduction: Many radiotracers have been used for sentinel node mapping with acceptable results. The main difference between these radiotracers is the particle size. In the current study, we reported defective labeling of Tc-99m antimony sulfide colloid which resulted in large particle size. Methods: Tc-99m-Antimony sulfide colloid was used for axillary sentinel node mapping of 45 breast cancer patients. The prepared kits were turbid and were used for the first 15 patients. For the remaining 30 patients, we used a filter (GyroDisc CA-PC Cellulose Acetate Membrane; 30 mm; Pore size: 0.2 μm) after labeling to remove the possible large particles of the prepared kits. Results: On the lymphoscintigraphy images, at least one sentinel node could be identified in 5 and 29 patients of the unfiltered and filtered groups respectively (p=0.00001). Sentinel node detection by gamma probe was successful in 5 and 30 patients in the unfiltered and filtered groups respectively (p=0.000001). Conclusion: Tc-99-Antimopny sulfide colloid is a suitable radiotracer for sentinel node mapping of the breast cancer patients. In case of any unusual turbidity of the labeled kit, it should not be used or at least be filtered before injection.
منابع مشابه
Sentinel node detection failure due to defective labeling and large particle size of Tc-99m antimony sulfide colloid
Introduction: Many radiotracers have been used for sentinel node mapping with acceptable results. The main difference between these radiotracers is the particle size. In the current study, we reported defective labeling of Tc-99m antimony sulfide colloid which resulted in large particle size. Methods: Tc-99m-Antimony sulfide colloid was used for axillary sentinel n...
متن کاملSentinel lymph node biopsy in melanoma patients: An experience with Tc-99m antimony sulfide colloid
Introduction: Sentinel lymph node biopsy is the standard procedure for lymph node staging in intermediate thickness melanoma. In Iran, this procedure has not been addressed sufficiently. In this study, we report our experience in this area. Methods: Ten consecutive patients with intermediate thickness melanoma where included in our study. 1.5 mCi of Tc-99m antimony...
متن کاملThe need for skin pen marking for sentinel lymph node biopsy: A comparative study
Introduction: There is a consensus in the literature that sentinel lymph node biopsy is the standard procedure for axillary staging in early stage (I and II) breast cancer patients. Usually during lymphoscintigraphy, the location of the sentinel lymph node is marked on the skin by an indelible ink. In this study we evaluated this issue in our patients. Methods: 40 ...
متن کاملParticle size analysis of (99m)Tc-labeled and unlabeled antimony trisulfide and rhenium sulfide colloids intended for lymphoscintigraphic application.
UNLABELLED Colloidal particle size is an important characteristic to consider when choosing a radiopharmaceutical for mapping sentinel nodes in lymphoscintigraphy. METHODS Photon correlation spectroscopy (PCS) and transmission electron microscopy (TEM) were used to determine the particle size of antimony trisulfide and rhenium sulfide colloids, and membrane filtration (MF) was used to determi...
متن کاملProduction of 99mTc-Antimony Sulfide Colloid for lymphoscintigraphy [Persian]
Introduction: The lymphatic system provides one of the main paths for the spread (Metastasis) of cancer from one part of body to another. Hodgkin’s diseases, lymphocytic leukemia, various metastatic disease and many lymph node disorders can be assessed by lymphoscintigraphy. Radionuclide lymphoscintigraphy has been used for many years to define lymphatic drainage of melanoma. The most com...
متن کامل