Five-year follow up of the ATS mechanical heart valve.

نویسندگان

  • Robert W Emery
  • Christopher C Krogh
  • Deb J Jones
  • Demetre M Nicoloff
  • David P Blake
  • Kit V Arom
چکیده

BACKGROUND AND AIM OF THE STUDY Between January 1, 1997 and December 31, 2001, a total of 342 patients underwent aortic valve replacement (AVR) or mitral valve replacement (MVR) with the ATS Medical prosthesis. The initial three-year phase of this study took place under a United States Food and Drug Administration-approved investigational device exemption study. The study aim was to determine the incidence of valve-related events in up to five years of follow up after valve implantation, and to assess patient disturbance from valve noise. METHODS Patients were consecutively enrolled to undergo AVR or MVR with the ATS prosthesis. Follow up studies were conducted by patient questionnaire and/or telephone call. Follow up was 96% complete. AVR was conducted in 246 patients (80 with coronary bypass), and MVR in 96 patients (29 with coronary bypass). RESULTS The overall operative mortality was 2.6% (n = 9; AVR 3.2%, n = 8; MVR 1.0%, n = 1), with two deaths being valve-related (0.6%). In 878 patient-years (pt-yr) of follow up (613 pt-yr for AVR; 265 pt-yr for MVR) there were an additional 22 deaths. Five deaths (0.6%/pt-yr) were valve-related: two were neuroembolic (both MVR), one from endocarditis (AVR), and two from bleeding events (both AVR). Late valve-related complications (>30 days) included 17 episodes of major bleeding (11 AVR, 1.8%/pt-yr; six MVR, 2.3%/pt-yr), five permanent neuroembolic events (four AVR, 0.7%/pt-yr; one MVR, 0.4%/pt-yr); 16 transient neuroembolic events (10 AVR, 1.6%/pt-yr; six MVR, 2.3%/pt-yr); three transient peripheral emboli (two AVR, 0.3%/pt-yr; one MVR, 0.4%/pt-yr); four paravalvular leaks (two AVR, 0.3%/pt-yr; two MVR, 0.8%/pt-yr); and one episode of valve thrombosis (MVR, 0.4%/pt-yr; AVR, 0%/pt-yr). Reoperation was required in two patients: one AVR (paravalvular leak, 0.2%/pt-yr) and one MVR (replacement due to thrombosis, 0.4%/pt-yr). CONCLUSION These results indicate that intermediate-term results with the ATS mechanical prosthesis continue to be excellent, though further long-term follow up is warranted.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Five-year follow-up after heart valve replacement with the CarboMedics bileaflet prosthesis.

BACKGROUND The CarboMedics valve is a relatively new, low-profile, bileaflet, mechanical prosthesis. The results of a prospective follow-up study after valve replacement with this prosthesis in a university hospital are presented. METHODS We implanted 640 CarboMedics prostheses in 583 patients in the aortic (n = 359), mitral (n = 167), or aortic and mitral positions (double valve replacement;...

متن کامل

Follow up the Treatment Process of Mitral Valve Diseases by Radiography and Echocardiography in Dogs

Background: Mitral valve disease is the most common acquired heart disease in dogs. Mitral valve regurgitation (MR) due to endocardiosis is an important cause of morbidity and mortality in dog. Echocardiography as a non-invasive tool is routinely used to evaluate and diagnose different disorders in the heart. Objectives: The purpose of this study is to determine the efficacy of medical treatmen...

متن کامل

Double Valve Replacement (Mitral and Aortic) for Rheumatic Heart Disease: A 20-year experience with 300 patients.

Introduction: Rheumatic heart disease still remains one of the leading causes of congestive heart failure and death owing to valvular pathologies, in developing countries. Valve replacement still remains the treatment of choice in such patients.The aim of this study wasto analyze the postoperative outcome of  double valve replacement (Mitral and Aortic ) in patients of rheumatic heart disease. ...

متن کامل

Prosthetic valve endocarditis caused by multidrug-resistant Candida albicans in a patient with myelodysplasia syndrome: A case report and literature review

Background and Purpose: Candida endocarditis is an infrequent disease with a high mortality rate, which commonly occurs in immunosuppressed patients with cardiac valve replacement. We reported a 70-year-old woman diagnosed with Candida prosthetic valve endocarditis (PVE). This study also involved a review of all published cases of Candida PVE from 1970. Case report: Herein, we reported a 70-ye...

متن کامل

Patient follow up care after open heart surgery.

OBJECTIVE: To determine the patient drop out to postoperative follow up for heart patients at the Kenyatta National Hospital. DESIGN SETTING AND PARTICIPANTS: The study analysed the postoperative attendance of heart patients to the surgical outpatient clinic at the KNH. Data on clinic attendance was collected over a one-year period from patient files and from clinic attendance. RESULTS: A total...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The Journal of heart valve disease

دوره 13 2  شماره 

صفحات  -

تاریخ انتشار 2004