Quality Comparison of Three Disposable Sperm Cell Counting Chambers
نویسندگان
چکیده
AND CONCLUSION In this study various quality aspects of the CellVision, Leja, and MicroCell disposable sperm cell counting slides were investigated. The height of the chambers was determined using interferometric spectrometry. The average heights of the Leja and MicroCell were very close, respectively 20.0 μm and 20.3 μm, whereas the CellVision was lower than specification (17.9 μm). The accuracy and reliability of concentration and motility measurements using the chambers was investigated by analyzing various semen samples with the Ceros CASA system of Hamilton Thorne. Counting with the Leja and MicroCell slides results in very similar concentrations, which are also very close to the FACS outcomes (used as external standard). However, the standard deviation and range of the Leja measurements are smaller than of MicroCell. This is probably caused by a non-homogeneous distribution of sperm cells within the MicroCell chamber; at the end of the chamber 70% above the average value was found, whereas deviations within the Leja slide do not exceed 7.4%. For these reasons, Leja is the most reliable chamber for concentration determination. The Leja slide has more advantages. First, less air bubbles are captured during chamber filling as compared to the MicroCell. Second, the highest motility values are measured in the Leja chamber. When using human sperm, motility in the MicroCell chamber is a few percent lower, but with the sensitive boar sperm, motilities in the MicroCell can be reduced by more than 50%. The CellVision chamber is extremely toxic for human, as well as for boar sperm. After 1 minute, hardly any motile cells can be detected anymore (for this reason, some experiments with the CellVision slide were omitted). The results of this study show that there is something to choose for.
منابع مشابه
Accuracy Evaluation of The Depth of Six Kinds of Sperm Counting Chambers for both Manual and Computer-Aided Semen Analyses
Background Although the depth of the counting chamber is an important factor influencing sperm counting, no research has yet been reported on the measurement and comparison of the depth of the chamber. We measured the exact depths of six kinds of sperm counting chambers and evaluated their accuracy. MaterialsAndMethods In this prospective study, the depths of six kinds of sperm counting chamber...
متن کاملAccuracy Evaluation of The Depth of Six Kinds of Sperm Counting Chambers for both Manual and Computer-Aided Semen Analyses.
BACKGROUND Although the depth of the counting chamber is an important factor influencing sperm counting, no research has yet been reported on the measurement and comparison of the depth of the chamber. We measured the exact depths of six kinds of sperm counting chambers and evaluated their accuracy. MATERIALS AND METHODS In this prospective study, the depths of six kinds of sperm counting cha...
متن کاملValidation of the Sperm Class Analyser CASA System for Sperm Count in a Diagnostic Semen Analysis Laboratory
Sperm counts have been linked to several fertility outcomes making it an essential parameter of semen analysis. It has become increasingly recognised that Computer Assisted Semen Analysis provides improved precision over manual methods but systems are seldom robustly validated for use. The objective of this work was to gather the evidence to validate or reject the Sperm Class Analyser as a tool...
متن کاملValidation of the sperm class analyser CASA system for sperm counting in a busy diagnostic semen analysis laboratory.
Sperm counts have been linked to several fertility outcomes making them an essential parameter of semen analysis. It has become increasingly recognised that Computer-Assisted Semen Analysis (CASA) provides improved precision over manual methods but that systems are seldom validated robustly for use. The objective of this study was to gather the evidence to validate or reject the Sperm Class Ana...
متن کاملMicro-cell chamber and washed human spermatozoa.
With the development of the Makler counting chamber, the reproducibility and accuracy of computer-assisted semen analysis and of manual determination of sperm concentration and sperm motility were significantly enhanced beyond what is possible with other methods, such as the hemocytometer, the Horwell chamber, the Petroff-Hausser chamber, and Chartpak slides. The use of Micro-Cell, a disposable...