Evaluation of the Diagnostic Performance of theSysmex XT-2000i Automated Hematology Analyzerin the Detection of Immature Granulocyte
نویسنده
چکیده
Immature Granulocytes (IG) are normally not present in peripheral blood. Their presence is an important indicator of enhanced bone marrow activation. The present study evaluated the diagnostic performance of the Sysmex XT-2000i automated hematology analyzer in the detection of myelocytes and/or metamyelocytes in patients affected by non-haematological disease, in comparison with manual microscopy review (NCCLS H20-A) and also in comparison with Sysmex XE-2100, used as a predicate device. Sysmex XT-2000i shows good specificity and sensitivity in the detection of low, medium and high concentrations of immature myeloid cells. The diagnostic performance of the XT-2000i IG and LS flags is very high and comparable to the results obtained with Sysmex XE-2100. These results suggest that Sysmex XT-2000i would be an excellent second analyser for larger laboratories concerned with harmonization of instrumentation (in particular as a back-up for Sysmex XE-2100), or as a primary instrument for medium-sized routine or STAT laboratories.
منابع مشابه
Performance characteristics of the Sysmex XT-2000i hematology analyzer.
The XT-2000i is a new hematology analyzer combining aperture impedance and optical technologies. It is capable of generating a 30-parameter hematologic assay on peripheral blood at a throughput of 80 samples per hour. We have evaluated the performance characteristics of the XT-2000i analyzer and compared them with those of the state-of-the-art XE-2100 analyzer and the manual differential count....
متن کاملPerformance evaluation of the immature granulocyte parameter on the Sysmex XE-2100 automated hematology analyzer.
Recently, St. Anthony's Medical Center Laboratory (SAMC) in St. Louis, Missouri, installed a new automated hematology analyzer, the Sysmex XE-2100. The XE-2100 has an additional differential parameter, the immature granulocyte (IG) parameter. Prior to implementation, the IG parameter was evaluated by comparison to the manual immature granulocyte count. Immature granulocytes counted manually, as...
متن کاملAutomated Hematology Analyzers in Diagnosis of Plasmodium vivax Malaria: an Adjunct to Conventional Microscopy
Malaria is one of the most pervasive parasitic diseases ever known to mankind affecting nearly 300 million people every year. The need for rapid diagnosis of malaria in tropical and subtropical malaria endemic areas is on the rise. In this study, we evaluated the usefulness of hematology autoanalyzers, Sysmex XE-2100 & XT-2000i in the presumptive diagnosis of malaria. Our study shows that abnor...
متن کاملAutomated Body fluid Analysis – A Comparative evaluation of diagnostic performance of Abotts Cell DYN Ruby &Sysmex -XT 2000i
If done meticulously , body fluid analysis has great diagnostic significance & implications . It can easily categorise the fluid into Inflammatory, infectious & malignant types. CSF analysis in bacterial meningitis is most dreaded & frequently encountered medical emergencies. CSF examination forms a important part to assess the prognosis, relapse & remission in Acute leukemias.Total WBC& differ...
متن کاملAutomated Measurement of Immature Granulocytes: Performance Characteristics and Utility in Routine Clinical Practice
The granulocytic "shift to left" reflects marrow response to bacterial infection, and this may be quantified as band count or immature granulocyte count (IGC). The former value, used widely in neonatal sepsis, has been notoriously difficult to measure accurately and precisely. A reproducible, precise, and accurate counting of immature granulocyte counts may be possible with automation. This stu...