Simple method for modulation transfer function determination of digital imaging detectors from edge images
نویسندگان
چکیده
A simple variant of the edge method to determine the presampled modulation transfer function (MTF) of digital imaging detectors has been developed that produces sufficiently accurate MTF values for frequencies up to the Nyquist frequency limit of the detector with only a small amount of effort for alignment and computing. An oversampled edge spread function (ESF) is generated from the image of a slanted edge by rearranging the pixel data of N consecutive lines that correspond to a lateral shift of the edge of one pixel. The original data are used for the computational analysis without further data preprocessing. Since the number of lines leading to an edge shift of one pixel is generally a fractional number rather than an integer, a systematic error may be introduced in the MTF obtained. Simulations and theoretical investigations show that for all frequencies up to the Nyquist limit the relative error ∆MTF/MTF is below 1/(2N) and can thus be kept below a given threshold by a suitable choice of N. The method is especially useful for applications where the MTF is needed for frequencies up to the Nyquist frequency limit, like the determination of the detective quantum efficiency (DQE).
منابع مشابه
محاسبۀ تابع انتقال مدولاسیون برای بررسی کنترل کیفی در رادیولوژی دیجیتال
The usage of digital radiology systems with insufficient quality control of digital image detectors can efface inherent superiorities of digital systems in comparison with analogue radiology systems and even makes an increased level of patients’ radiation doses. Modulation transfer function (MTF) is one of the most important parameters related to image quality and digital detectors efficiency. ...
متن کاملEvaluation of the Portal Imaging System Performance for an Elekta Precise Linac in Radiotherapy
Introduction: Electronic portal imaging devices (EPIDs) provide two- and three-dimensional planar and volumetric cone beam images to improve the accuracy of radiation treatment delivery. Periodic quality assurance (QA) of EPIDs is essential for dosimetric verification in radiotherapy. In this study, a QA program was implemented to evaluate the function of the EPID to be confident in applying co...
متن کاملDesign and Fabrication Process of MTF Phantom CT Scan
Introduction: One of the main steps in the optimization process in diagnostic imaging is the quality control of radiology devices. The usual method of CT scan calibration is used of a phantom. The phantom created a certain weakening for the radiation through which it passes. One of the most suitable methods for quantitative analysis of the resolution and contrast in CT scan im...
متن کاملAlgorithmic Analysis of Edge Ranking and Profiling for MTF Determination of an Imaging System
Edge detection is one of the most principal techniques for detecting discontinuities in the gray levels of image pixels. The Modulation Transfer Function (MTF) is one of the main criteria for assessing imaging quality and is a parameter frequently used for measuring the sharpness of an imaging system. In order to determine the MTF, it is essential to determine the best edge from the target imag...
متن کاملA Proposed Simple Teleradiology System Using a Digital Camera
Background & Aims: Teleradiology as an efficient way for avoiding unnecessary transfer of patients to special medical centers requires the images with digital format. However, in some developing countries, including Iran, imaging is being done using non digital devices. In such cases, changing analog images into digital ones can solve the problem. The method which is nowadays used for digita...
متن کامل