Factors Related to Radiation Exposure during Lumbar Spine Intervention
نویسندگان
چکیده
Fluoroscopy guidance is useful to confirm anatomical landmark and needle location for spine intervention; however, it can lead to radiation exposure in patients, physicians, and medical staff. Physicians who used fluoroscopy should be cognizant of radiation exposure and intend to minimize radiation dose. We retrospectively reviewed three lumbar spine intervention procedures (nerve root block, medial branch block, and facet joint block) at our institution between June and December, 2014. We performed 268 procedures on 220 patients and found significant difference in radiation dose between two groups classified by performing physicians. The physician who controlled the fluoroscopy unit directly used significantly shorter fluoroscopy (6 seconds) that resulted in a smaller radiation dose (dose area product [DAP] 0.59 Gy∙cm(2)) than the physician supervising the radiographer controlling the fluoroscopy unit (72 seconds, DAP 5.31 Gy∙cm(2), P < 0.001). The analysis indicates that the difference in fluoroscopy time depends on whether a physician or a radiographer controls the fluoroscopy unit.
منابع مشابه
Beam Collimation during Lumbar Spine Radiography: A Retrospective Study
Introduction: Collimating the primary beam to the area of diagnostic interest (ADI) has been strongly recommended as an effective method to reduce patient’s radiation dose and to improve image quality during radiology practice. Lack or inadequate collimation results in excessive radiation dose to patients and deterioration image quality.Objective: To assess the quality of beam collimation dur...
متن کاملارزیابی دز ورودی پوستی و دز مؤثر بیماران در معمول ترین آزمون های رادیولوژی
Introduction: Nowadays, we are witnessing a progressive application of radiology techniques in various communities as well as in Iran. Medical radiological examinations are one of the important sources of public radiological dose. Entrance skin doses and effective doses for patients in diagnostic radiology examinations should be assessed for the imaging techniques optimization and patient&rsquo...
متن کاملEntrance skin dose assessment of selected computed radiography facilities in Ghana
Background: The basic challenge with computed radiography (CR) systems is the large dynamic range which provides an opportunity for radiographers to gradually increase exposure factors and still produce good image quality, a practice that can lead to dose creep. Materials and Methods: The aim of this study was to establish the entrance skin dose (ESD) values for nine selected examinations in t...
متن کاملRadiation exposure during the lateral lumbar interbody fusion procedure and techniques to reduce radiation dosage
BACKGROUND Fluoroscopy is widely used in spine surgery to assist with graft and hardware placement. Previous studies have not measured radiation exposure to a surgeon during minimally invasive lateral lumbar spine surgery for single-level discectomy and interbody cage insertion. This study was performed to model and measure radiation exposure to a surgeon during spine surgery using the direct l...
متن کاملEffective radiation dose reduction in computed tomography-guided spinal injections: a prospective, comparative study with technical considerations
Despite the good general patient acceptance, high patient comfort, safety and precision in the needle placement, exposure to radiation in computed tomography (CT)-guided spinal interventions remains a serious concern, and is often used to argue against its use. The aim of this study was to determine the technical possibilities of reducing the radiation dose in CT-guided epidural and periradicul...