Rupture of Accura balloon during percutaneous transvenous mitral commissurotomy
نویسنده
چکیده
Percutaneous transvenous mitral commissurotomy (PTMC) is one of the treatment modality in the management of rheumatic mitral stenosis. Reuse of sterilized PTMC balloon catheters has been widely practiced in developing countries to bring down procedural cost. PTMC by reused balloons are prone to deform and rupture during procedure. This can cause fatal complications like embolism of air or balloon material. Herein, we report a case of Accura balloon rupture during PTMC without any fatal complication. Thus, during balloon preparation particularly of reused balloon, it should be examined for any deformity or tear and air should be removed completely to prevent the fatal outcome.
منابع مشابه
The Association between B-type Natriuretic Peptide Family and Successful Percutaneous Transvenous Mitral Commissurotomy: A Systematic Review
B-type natriuretic peptide (BNP) level is known to increase in patients with rheumatic mitral stenosis. In this systematic review, we aimed to discuss the possible association between plasma BNP level and the success rate of percutaneous transvenous mitral commissurotomy. PubMed and Scopus databases were searched systematically, using the following key terms: “B-type natriuretic peptide” OR “BN...
متن کاملSHORT TERM CLINICAL OUTCOME OF PERCUTANEOUS BALLOON MITRAL VALVULOPLASTY VERSUS SURGICAL CLOSED MITRAL COMMISSUROTOMY
BaIlon mitral valvuloplasty and closed mitral valve conunissurotomy were performed in 450 and 127 patients, respectively. Also, 254 patients with severe mitral stenosis were enrolled in a prospective randomized trial comparing the two procedures. Immediately after balloon mitral valvuloplasty, mean left atrial pressure was 11.38±3.54 mmHg, transmitral valve gradient was 1.8±2 mmHg, and mit...
متن کاملAtrial Septa1 Puncture Technique in Percutaneous Transvenous Mitral Commissurotomy
Transseptal catheterization is a vital component of percutaneous transvenous mitral commissurotomy. Therefore, a well-executed transseptal catheterization is the key to a safe and successful percutaneous transvenous mitral commissurotomy. Two major problems inherent in atrial septel puncture for percutaneous transvenous mitral cornmissurotomy are cardiac perforation and puncture of an inappropr...
متن کاملPercutaneous transvenous mitral commissurotomy in juvenile mitral stenosis.
OBJECTIVE To determine the efficacy and safety of percutaneous transvenous mitral commissurotomy (PTMC), using multi-track double balloon technique in juvenile mitral stenosis. DESIGN Open non-randomised intervention. SETTING Cardiac catheterisation laboratories of The Mater Hospital, The Nairobi Hospital and Kenyatta National Hospital from 1996 to 2001. PATIENTS Forty five consecutive pa...
متن کاملDeflation failure of Accura Balloon In Left Atrium During Balloon Mitral Valvuloplasty
Currently Percutaneous Balloon Mitral Valvuloplasty has established itself as a standard treatment for patients with severe mitral stenosis. A patient with severe mitral stenosis was posted for the above procedure. After dilatation of mitral valvular apparatus the Accura balloon failed to deflate on repeated aspirations, probably due to rupture of internal lumen. Finally, a guide wire was passe...
متن کامل